Reliable Acquisition of Electroencephalography Data during Simultaneous Electroencephalography and Functional MRI
Published on: March 19, 2021
Soft, adhesive and conductive composite for electroencephalogram signal quality improvement
Jeong E Jin1, Seohyeon Kim1, Hyeji Yu1
1School of Electrical Engineering, Kookmin University, Seoul, 02707 South Korea.
Abstract:
Since electroencephalogram (EEG) is a very small electrical signal from the brain, it is very vulnerable to external noise or motion artifact, making it difficult to measure. Therefore, despite the excellent convenience of dry electrodes, wet electrodes have been used. To solve this problem, self-adhesive and conductive composites using carbon nanotubes (CNTs) in adhesive polydimethylsiloxane (aPDMS), which can have the advantages of both dry and wet electrodes, have been developed by mixing them uniformly with methyl group-terminated PDMS. The CNT/aPDMS composite has a low Young's modulus, penetrates the skin well, has a high contact area, and excellent adhesion and conductivity, so the signal quality is enhanced. As a result of the EEG measurement test, although it was a dry electrode, results comparable to those of a wet electrode were obtained in terms of impedance and motion noise. It also shows excellent biocompatibility in a human fibroblast cell test and a week-long skin reaction test, so it can measure EEG with high signal quality for a long period of time.