Operating room disinfection: operator-driven ultraviolet 'C' vs. chemical treatment
Marie-Claire Fickenscher1, Madeline Stewart1, Ryan Helber1
1Hoag Orthopedics Education and Research, Hoag Orthopedic Institute, Irvine, CA, United States.
Background:
In operating room (OR) surfaces, Nosocomial pathogens can persist on inanimate surfaces for long intervals and are highly resistant to traditional surface cleaning.
Aim:
This study compares traditional chemical operating room terminal disinfection to a unique operator-driven device that emits germicidal UV light at short distance onto vertical and horizontal surfaces.
Methods:
A randomized crossover analogous protocol assigned 40 end-of-day operating rooms into either group A (chemical then UVC treatments) or group B (UVC then chemical treatments). Initial Staphylococcal cultures were obtained prior to disinfection treatment, after the first treatment, and after the second treatment at 16 most commonly contaminated sites to represent overall room contamination. Success was defined as no growth and failure as 1 or more colony forming units. Thoroughness of chemical treatment vs UVC treatment was compared and used to determine if the second treatment was additive to the first treatment within each group.
Findings:
The operator driven UVC device outperformed chemical treatment in reducing the number of contaminated sites in the OR by more than half (P<0.001). Operator-driven UVC reduced contaminated sites after chemical treatment by nearly half (P<0.001). In contrast, chemical treatment after operator-driven UVC did not significantly reduce the number of contaminated sites. The mean employee time of disinfection for chemical treatment was 49 minutes and for the operator-driven UVC emitter 7.9 minutes (P<0.001).
Conclusions:
This study demonstrates that addition of an operator-driven UVC emitter to OR rooms between cases could be helpful in overall decreasing the number of contaminated sites.
Related Concept Videos
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Chemical Agents for Microbial Control
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Hand hygiene
Hand washing...
Physical Methods for Controlling Microbial Growth: Radiation and Filtration
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...