Evaluation of Compressed SENSE on Image Quality and Reduction of MRI Acquisition Time: A Clinical Validation Study
Pohchoo Seow1, Si Wei Kheok1, Meng Ai Png1
1Department of Diagnostic Radiology, Singapore General Hospital, 31, Third Hospital Ave, Central Region, Singapore 168753; Duke-NUS Medical School, Central Region, Singapore.
Rationale And Objectives:
To evaluate the effect of compressed SENSE (CS) in clinical settings on scan time reduction and image quality.
Materials And Methods:
Ninety-five magnetic resonance imaging (MRI) scans from different anatomical regions were acquired, consisting of a standard protocol sequence (SS) and sequence accelerated with CS. Anonymized paired sequences were randomly displayed and rated by six blinded subspecialty radiologists. Side-by-side evaluation on perceived sharpness, perceived signal-to-noise-ratio (SNR), lesion conspicuity, and artifacts were compared and scored on a five-point Likert scale, and individual image quality was evaluated on a four-point Likert scale.
Results:
CS reduced overall scan time by 32% while maintaining acceptable MRI quality for all regions. The largest time savings were seen in the spine (mean = 68 seconds, 44% reduction) followed by the brain (mean = 86 seconds, 37% reduction). The sequence with maximum time savings was intracranial 3D-time-of-flight magnetic resonance angiography (202 seconds, 56% reduction). CS was mildly inferior to SS on perceived sharpness, perceived SNR, and lesion conspicuity (mean scores = 2.32-2.96, P < .001 [1: SS superior; 3: equivalent; 5: CS superior]). CS was equivalent to SS for joint and body scans on overall image quality (CS = 3.02-3.37, SS = 3.04-3.68, P > .05, [1: lowest quality and 4: highest quality]). The overall image quality of CS was slightly less for brain and spine scans (mean CS = 2.79-3.05, mean SS = 3.13-3.43, P = .021) but still diagnostic. Good overall clinical acceptance for CS (88%) was noted with full clinical acceptance for body scans (100%) and high acceptance for other regions (68%-95%).
Conclusion:
CS significantly reduced MR acquisition time while maintaining acceptable image quality. The implementation of CS may improve departmental workflows and enhance patient care.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System IV: CMRI
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies IV: Magnetic Resonance Imaging