Comparison of ASTM F2129 and ASTM F746 for Evaluating Crevice Corrosion
Shiril Sivan1, Elnaz Rahman1, Jason D Weaver1
1Division of Applied Mechanics, Office of Science and Engineering Laboratories, Center for Devices and Radiological Health, U.S Food and Drug Administration, 10903 New Hampshire Ave., Silver Spring, MD 20993, USA.
None:
Crevice corrosion is one of the major mechanisms that drives implant failure in orthopedic devices that have modular interfaces. Despite the prevalence of crevice corrosion in modular interfaces, very little is known with regards to the susceptibility of different material combinations to participate in crevice corrosion. In this study, we compare two electrochemical methods, ASTM F2129, Standard Test Method for Conducting Cyclic Potentiodynamic Polarization Measurements to Determine the Corrosion Susceptibility of Small Implant Devices, and a modified version of ASTM F746, Standard Test Method for Pitting or Crevice Corrosion of Metallic Surgical Implant Materials, in their ability to induce crevice corrosion. Four commonly used metals, 316 stainless steel, commercially pure titanium (Ti grade 2), Ti-6Al-4V (Ti grade 5), and cobalt-chromium-molybdenum per ASTM F1537, Standard Specification for Wrought Cobalt-28Chromium-6Molybdenum Alloys for Surgical Implants (UNSR31537, UNSR31538, and UNSR31539), were used to form crevices with a rod and washer combination. As a control, the metal rod materials were tested alone in the absence of crevices using ASTM F2129 and the modified ASTM F746 method. As another control to determine if crevices formed with polymeric materials would influence crevice corrosion susceptibility, experiments were also conducted with metal rods and polytetrafluorethylene washers. Our results revealed more visible corrosion after ASTM F2129 than ASTM F746. Additionally, ASTM F746 was found to falsely identify crevice corrosion per the critical pitting potential when visual inspection found no evidence of crevice corrosion. Hence, ASTM F2129 was found to be more effective overall at evaluating crevice corrosion compared to ASTM F746.
More Related Videos
10:00Experimental Protocol to Determine the Chloride Threshold Value for Corrosion in Samples Taken from Reinforced Concrete Structures
Published on: August 31, 2017
07:37Full-field Strain Measurements for Microstructurally Small Fatigue Crack Propagation Using Digital Image Correlation Method
Published on: January 16, 2019
Related Concept Videos
Testing Water Quality
Corrosion of Reinforcement
However, over time and under certain conditions like carbonation, chloride ingress, and cracking this protective state can be compromised. Steel has areas with...
Strength of Cement
For compressive strength tests, ASTM C 109-05 standards prescribe a cement-sand mix ratio of 1:2.75 and a water/cement ratio of 0.485 for making 2-inch cubes. These cubes are mixed, cast, and cured in saturated lime water at 23°C until testing. Flexural strength testing, outlined in...
Mechanical Characteristics of Steel
The tension test is fundamental for determining tensile strength. In this test, a steel specimen is stretched using a gripping device until it breaks. The data collected during this test are used...
Deleterious Substances in Aggregate
Another type of impurity is clay and fine material that...
Microcracking in Concrete