Related Experiment Video
Updated: Jul 16, 2025

10:10
Reverse Total Shoulder Arthroplasty
Published on: July 5, 2011
43.3K
Reconstruction with a double-constrained implant design after complex shoulder extra-articular resection
Jan Lesensky1, Ana C Belzarena2, Matej Daniel3
1Department of Orthopaedics, First Medical Faculty, University Hospital Na Bulovce, Charles University, Prague, Czech Republic. janless@gmail.com.
World Journal of Surgical Oncology
|September 18, 2023
Summary
A novel constrained shoulder implant offers a promising solution for patients undergoing extra-articular resection of proximal humerus tumors. This innovative design reduces stress on the glenoid component, improving stability and patient outcomes.
Area of Science:
- Orthopedic surgery
- Biomedical engineering
- Oncology
Background:
- Extra-articular resection for proximal humerus tumors often leads to poor functional outcomes and high revision rates.
- Existing implants face challenges: unconstrained types cause instability, while constrained ones risk glenoid component loosening due to bone overload.
Purpose of the Study:
- To develop a constrained implant for extra-articular resection cases with deltoid and rotator cuff function loss.
- The implant aims to provide stability, passive motion, and reduce glenoid component aseptic loosening.
Main Methods:
- A novel implant with two constrained joints in series, connected by a dumbbell piece, was designed in collaboration with Czech Technical University.
- Biomechanical analysis demonstrated significant load reduction on the glenoid component.
- Three patients with extensive tumor resection received the implant, with an average follow-up of 16 months.
Main Results:
- The surgical technique was straightforward, with an average surgery time of 175 minutes and 516 ml blood loss.
- No surgical or implant-related complications occurred in the three patients.
- Patients reported pain-free, stable shoulders, fully functional extremities, and high satisfaction, with an average MSTS score of 70%.
Conclusions:
- The innovative constrained implant design shows potential as an effective reconstruction method for challenging proximal humerus tumor resections.
- This approach addresses limitations of previous implants, offering improved stability and reduced risk of loosening.