Effect of shape of automatic dose rate control and wedge compensation filter on radiation dose in an angiography
Kazuya Kakuta1,2, Shumpei Nemoto3, Masamitsu Ikeda3
1Department of Disaster Medicine, Fukushima Medical University Hospital, Fukushima, Japan. kakuta-k@fmu.ac.jp.
Abstract:
The purpose was to investigate air-kerma area product (PKA) and entrance surface air-kerma rate ([Formula: see text]a,e) on the effect of the shape of automatic dose rate control (ADRC) in the presence of a wedge compensation filter. We compared and evaluated the variability of the X-ray output using a combination of wedge compensation filters and the ADRC. Two ADRC shapes (round and square) and three poly-methyl-methacrylate thicknesses (15, 20, and 25 cm) were used. A wedge compensation filter was inserted 2 cm at a time, up to 6 cm. When the wedge compensation filter was inserted to 6 cm for 20 cm of poly-methyl-methacrylate, the X-ray output fluctuated significantly. The PKA was reduced by 39% when the wedge compensation filter was inserted to 6 cm and by 59% when it was inserted to 4 cm under round-type for 20 cm poly-methyl-methacrylate. The shape of the ADRC affects [Formula: see text]a,e and PKA.