The first investigation of spot-scanning proton arc (SPArc) delivery time and accuracy with different delivery
Gang Liu1,2,3, Lewei Zhao4,5, Peilin Liu4
1Cancer Center, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan 430022, People's Republic of China.
Abstract:
Objective. To investigate the impact of various delivery tolerance window settings on the treatment delivery time and dosimetric accuracy of spot-scanning proton arc (SPArc) therapy.Approach. SPArc plans were generated for three representative disease sites (brain, lung, and liver cancer) with an angle sampling frequency of 2.5°. An in-house dynamic arc controller was used to simulate the arc treatment delivery with various tolerance windows (±0.25, ±0.5, ±1, and ±1.25°). The controller generates virtual logfiles during the arc delivery simulation, such as gantry speed, acceleration and deceleration, spot position, and delivery sequence, similar to machine logfiles. The virtual logfile was then imported to the treatment planning system to reconstruct the delivered dose distribution and compare it to the initial SPArc nominal plan. A three-dimensional gamma index was used to quantitatively assess delivery accuracy. Total treatment delivery time and relative lost time (dynamic arc delivery time-fix beam delivery time)/fix beam delivery time) were reported.Main Results. The 3D gamma passing rate (GPR) was greater than 99% for all cases when using 3%/3 mm and 2%/2 mm criteria and the GPR (1%/1 mm criteria) degraded as the tolerance window opens. The total delivery time for dynamic arc delivery increased with the decreasing delivery tolerance window length. The average delivery time and the relative lost time (%) were 630 ± 212 s (253% ± 68%), 322 ± 101 s (81% ± 31%), 225 ± 60 s (27% ± 16%), 196 ± 41 s (11% ± 6%), 187 ± 29 s (6% ± 1%) for tolerance windows ±0.25, ±0.5, ±1, and ±1.25° respectively.Significance. The study quantitatively analyzed the dynamic SPArc delivery time and accuracy with different delivery tolerance window settings, which offer a critical reference in the future SPArc plan optimization and delivery controller design.