Quantitative 177Lu SPECT/CT imaging for personalized dosimetry using a ring-shaped CZT-based camera.
Rachele Danieli1,2, Martina Stella3,4, Julian Leube5
1Department of Medical Physics, Institut Jules Bordet, Hôpital Universitaire de Bruxelles (H.U.B), Université Libre de Bruxelles (ULB), Brussels, Belgium. rachele.danieli@hubruxelles.be.
Accurate dosimetry for lutetium-177 radiopharmaceutical therapy (177Lu-RPT) is achievable with StarGuide™ SPECT/CT imaging. Both OSEM and BSREM reconstruction algorithms offer comparable performance, with careful parameter selection crucial for optimal results in patient dosimetry.
Area of Science:
- Nuclear Medicine
- Medical Imaging
- Radiopharmaceutical Therapy
Background:
- Accurate dosimetry is essential for radiopharmaceutical therapy (177Lu-RPT) using quantitative SPECT/CT imaging.
- Developing suitable reconstruction protocols is critical for reliable SPECT/CT quantitative performance.
- This study evaluates a CZT-based camera for 177Lu imaging.
Purpose of the Study:
- To characterize the quantitative performance of a ring-shaped CZT-based camera for 177Lu imaging.
- To compare two reconstruction algorithms: OSEM and BSREM (Q.Clear) for 177Lu dosimetry.
- To establish foundations for defining a reconstruction protocol for accurate 177Lu-RPT dosimetry.
Main Methods:
- Acquired 177Lu-filled phantoms using a StarGuide™ camera.
- Reconstructed images using manufacturer implementations of OSEM (GE-OSEM) and BSREM (Q.Clear) with varying iterations and subsets.
- Evaluated quantification accuracy, noise, recovery coefficients (RCs), and spatial resolution in phantoms and patient cohorts.
Main Results:
- Activity accuracy was within 10% for all reconstructions, despite septal penetration.
- Both OSEM and BSREM improved spatial resolution and RCs with increased updates, but BSREM showed better noise reduction.
- Carefully selected parameters yielded comparable noise, spatial resolution (~1 cm), and RCs across algorithms in phantoms and patients.
Conclusions:
- Accurate 177Lu quantification is feasible with StarGuide™ SPECT/CT.
- GE-OSEM provides a valid alternative to Q.Clear, demonstrating comparable performance for phantom and patient dosimetry.