Related Experiment Video
Updated: Jul 11, 2025

Measuring Biophysical and Psychological Stress Levels Following Visitation to Three Locations with Differing Levels of Nature
Published on: June 19, 2019
Diagnostic performance of an automated immunoassay for salivary cortisol
Christelle Laguillier-Morizot1, Fidéline Bonnet-Serrano1, Marie-Clémence Leguy2
1Department of Hormonology, bâtiment J. Dausset, 2(e) étage, CHU Cochin, 27, rue du Fg-St-Jacques, 75014 Paris, France; Faculté de santé, université Paris Cité, 2, rue Valette, 75005 Paris, France.
Abstract:
Automated immunoanalysis (AI) is an interesting alternative for measuring salivary cortisol, as the gold standard HPLC-MS/MS method is not yet readily available. The aim of this study was to evaluate the diagnostic performance of salivary cortisol immunoassay on the iSYS immunoanalyzer in adrenal dynamic tests. Cortisol was measured on iSYS and on HPLC-MS/MS in saliva samples collected after 1mg-dexamethasone suppression test (DST) in 115 patients suspected of Cushing syndrome, and during Synacthen® stimulation test (SST) in 108 patients suspected of adrenal insufficiency. Concentrations on AI correlated well with HPLC-MS/MS (Spearman r=0.9496; P<0.0001), but with a significant positive bias. ROC analysis of salivary cortisol identified optimal cut-off values on AI and HPLC-MS/MS of respectively 3.5 and 0.77nmol/L for DST and 32.6 and 13.8nmol/L at T60 after SST. Automated immunoassays for salivary cortisol are suitable in daily practice but require determination of specific cut-off and reference values.

