Related Experiment Video
Updated: Jul 11, 2025

Use of Rabbit Eyes in Pharmacokinetic Studies of Intraocular Drugs
Published on: July 23, 2016
Fluocinolone Acetonide Implant Removal From the Vitreous Cavity
Nita Valikodath1, Lejla Vajzovic2
1Department of Ophthalmology, University of Michigan, Ann Arbor, MI, USA.
Abstract:
Purpose: To describe a technique for fluocinolone acetonide implant removal from the vitreous cavity. Methods: A case report and review of surgical methods were performed. Results: The technique to remove a fluocinolone acetonide implant from the vitreous cavity was safe and effective. The vitreous cutter was used to align the implant coaxially with a 25-gauge cannula. The valve of the cannula was opened, creating pressure that drew the implant into the cannula. Conclusions: Fluocinolone acetonide implants can be efficiently removed from the vitreous cavity by creating a pressure differential using the valved cannula. Advantages of this technique include avoiding invasive maneuvers such as enlarging the sclerotomy or creating a corneal wound.
More Related Videos
11:20Retinal Pigment Epithelium Transplantation in a Non-human Primate Model for Degenerative Retinal Diseases
Published on: June 14, 2021
04:59Author Spotlight: Enhancing Visual Outcomes in Cataract Surgery: A Novel Technique to Prevent Posterior Capsular Opacification Through IOL Rotation
Published on: July 7, 2023