Organ-contour-driven auto-matching algorithm in image-guided radiotherapy
Yukako Kishigami1, Mitsuhiro Nakamura1, Hiroyuki Okamoto2
1Department of Advanced Medical Physics, Graduate School of Medicine, Kyoto University, Kyoto, Japan.
Journal of Applied Clinical Medical Physics
|November 23, 2023
Summary
An organ-contour-driven auto-matching algorithm for image-guided radiotherapy offers clinical potential. This method ensures consistent positioning and minimal impact from rotational corrections on planning target volume coverage.
Area of Science:
- Medical Physics
- Radiation Oncology
- Image Processing
Background:
- Image-guided radiotherapy (IGRT) requires precise patient positioning.
- Organ contour information can enhance the accuracy of IGRT positioning.
- Cervical cancer treatment involves complex target volumes and organs at risk.
Purpose of the Study:
- To evaluate the clinical applicability of an organ-contour-driven auto-matching algorithm for IGRT.
- To assess the algorithm's ability to generate consistent and accurate patient positioning.
- To determine the impact of rotational corrections on planning target volume (PTV) coverage.
Main Methods:
- Developed a weight-based auto-matching algorithm using mesh models of daily and reference MRI scans.
- Implemented a cost function prioritizing specific organs for positioning in three scenarios.
- Optimized translation and rotation values for cervix and uterus, with a rotation limit of ±3°.
- Evaluated PTV coverage using isotropic 5 mm and anisotropic margins.
Main Results:
- Translations exceeding 10 mm occurred in 9-18% of fractions, varying by scenario.
- Over 80% of fractions achieved ≥80% coverage for small PTVs, and >90% achieved ≥95% coverage for large PTVs.
- Rotational corrections (±3°) had a minimal impact (<1% difference in median coverage) compared to translational positioning alone.
Conclusions:
- The organ-contour-driven algorithm enables quantitative and prioritized organ positioning in IGRT.
- Consistent displacement values were algorithmically generated, demonstrating reliability.
- Limited rotational corrections (±3°) minimally affected PTV coverage, simplifying the positioning process.