Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function
Published on: December 11, 2019
Diagnostic Accuracy of Single-Lead Electrocardiograms Using the Kardia Mobile App and the Apple Watch 4: Validation
Kristina Klier1, Lucas Koch1, Lisa Graf1
1Institute of Sport Science, University of the Bundeswehr Munich, Neubiberg, Germany.
Background:
To date, the 12-lead electrocardiogram (ECG) is the gold standard for cardiological diagnosis in clinical settings. With the advancements in technology, a growing number of smartphone apps and gadgets for recording, visualizing, and evaluating physical performance as well as health data is available. Although this new smart technology is innovative and time- and cost-efficient, less is known about its diagnostic accuracy and reliability.
Objective:
This study aimed to examine the agreement between the mobile single-lead ECG measurements of the Kardia Mobile App and the Apple Watch 4 compared to the 12-lead gold standard ECG in healthy adults under laboratory conditions. Furthermore, it assessed whether the measurement error of the devices increases with an increasing heart rate.
Methods:
This study was designed as a prospective quasi-experimental 1-sample measurement, in which no randomization of the sampling was carried out. In total, ECGs at rest from 81 participants (average age 24.89, SD 8.58 years; n=58, 72% male) were recorded and statistically analyzed. Bland-Altman plots were created to graphically illustrate measurement differences. To analyze the agreement between the single-lead ECGs and the 12-lead ECG, Pearson correlation coefficient (r) and Lin concordance correlation coefficient (CCCLin) were calculated.
Results:
The results showed a higher agreement for the Apple Watch (mean deviation QT: 6.85%; QT interval corrected for heart rate using Fridericia formula [QTcF]: 7.43%) than Kardia Mobile (mean deviation QT: 9.53%; QTcF: 9.78%) even if both tend to underestimate QT and QTcF intervals. For Kardia Mobile, the QT and QTcF intervals correlated significantly with the gold standard (rQT=0.857 and rQTcF=0.727; P<.001). CCCLin corresponded to an almost complete heuristic agreement for the QT interval (0.835), whereas the QTcF interval was in the range of strong agreement (0.682). Further, for the Apple Watch, Pearson correlations were highly significant and in the range of a large effect (rQT=0.793 and rQTcF=0.649; P<.001). CCCLin corresponded to a strong heuristic agreement for both the QT (0.779) and QTcF (0.615) intervals. A small negative correlation between the measurement error and increasing heart rate could be found of each the devices and the reference.
Conclusions:
Smart technology seems to be a promising and reliable approach for nonclinical health monitoring. Further research is needed to broaden the evidence regarding its validity and usability in different target groups.
More Related Videos
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage....