Electrochemical detection of creatinine: exploiting copper(ii) complexes at Pt microelectrode arrays
Keerakit Kaewket1, Kamonwad Ngamchuea1
1School of Chemistry, Institute of Science, Suranaree University of Technology 111 University Avenue, Suranaree, Muang Nakhon Ratchasima 30000 Thailand kamonwad@g.sut.ac.th +66 (0) 44 224 637.
Abstract:
This work develops a rapid and highly sensitive electrochemical sensor for creatinine detection at platinum microelectrode arrays (Pt-MEA). Copper(ii) ions are introduced to form the electroactive creatinine complex, which is then detected at Pt-MEA through a direct reduction reaction. Electrochemical behaviors of the creatinine complex are also explored at Pt macrodisc and microdisc electrodes in comparison with Pt-MEA. At the Pt-MEA, the linear range, sensitivity, and limit of detection of creatinine are determined to be 0.00-5.00 mM, 5401 ± 99 A m-2 M-1, and 0.059 mM (3SB/m), respectively. Notably, the Pt-MEA requires only 10 μL of sample and allows direct measurement of creatinine in synthetic urine with 97.39 ± 4.78% recovery.