Automated magnetic-bead-assisted sequential extraction technology for simultaneous detection of Aβ1-42 and Aβ1-40 in cerebrospinal fluid: An advance toward fully automated liquid chromatography-tandem mass spectrometry method
Related Experiment Video
Updated: Jul 9, 2025

Absolute Quantification of Aβ1-42 in CSF Using a Mass Spectrometric Reference Measurement Procedure
Published on: March 21, 2017
Yutong Zou1, Xiaoli Ma1, Chenhui Mao2
1Department of Laboratory Medicine, Peking Union Medical College & Chinese Academy of Medical Sciences, Peking Union Medical College Hospital, Beijing 100730, PR China.
Abstract:
Traditional solid-phase extraction (SPE) LC-MS/MS is limited by high costs, turnaround times, and procedural complexity, which limited the usage in clinical practice. This study aimed to establish a robust UPLC-MS/MS method with automated magnetic-bead-assisted sequential extraction (MBASE) technology to simultaneously measure Aβ1-42 and Aβ1-40 in cerebrospinal fluid (CSF). A Waters TQ-XS triple quadrupole mass spectrometer and Acquity UPLC Protein BEH C4 column were used. The targeted analytes were extracted and concentrated using the automated MBASE technology with chemically modified magnetic MCX beads. Analytical performance was verified referring to the CLSI C62-A and EP-15-A3 guidelines. A total of 68 CSF samples were collected and analyzed using the MBASE UPLC-MS/MS method, traditional SPE UPLC-MS/MS method, and Lumipulse G fully automated chemiluminescence detection system, and method comparison analysis is conducted. The MBASE UHPLC-MS/MS method showed an analytical performance equivalent to that of traditional SPE technology, with a higher sample throughput and smaller amount of materials ($34.98 vs. $493.96) and labor cost (101 min vs. 140 min) for 96 samples. The limit of quantification (LOQ) of Aβ1-42 and Aβ1-40 was 0.10 ng/mL and 0.05 ng/mL; recovery was 88.35-107.07 % and 95.72-96.60 %; and total imprecision was 3.69-6.83 % and 3.02-3.61 %, respectively. The measurements were faithfully reproduced within the allowable levels of uncertainty using certified reference materials. The correlations between this MBASE UPLC-MS/MS method, the SPE UPLC-MS/MS method, and Lumipulse G fully automated biochemical analysis method are all deemed good (r = 0.869-0.936), and the MBASE- and SPE-UPLC-MS/MS methods showed comparable measurements. To our knowledge, our study firstly verified the robust performance of the MBASE UPLC-MS/MS method to simultaneously determine Aβ1-42 and Aβ1-40 in CSF. With further introduce of automation, the assay with high accuracy and low material and labor costs will become a promising clinical technology.
More Related Videos
07:08A High Throughput, Multiplexed and Targeted Proteomic CSF Assay to Quantify Neurodegenerative Biomarkers and Apolipoprotein E Isoforms Status
Published on: October 20, 2016
11:03Use of Capillary Electrophoresis Immunoassay to Search for Potential Biomarkers of Amyotrophic Lateral Sclerosis in Human Platelets
Published on: February 10, 2020