Related Experiment Video
Updated: Jul 9, 2025

Noninvasive In Vivo Small Animal MRI and MRS: Basic Experimental Procedures
Published on: October 20, 2009
Inductively coupled, transmit-receive coils for proton MRI and X-nucleus MRI/MRS in small animals
Atsushi M Takahashi1, Jitendra Sharma1,2,3,4, David O Guarin5,2
1Department of Brain and Cognitive Sciences, Massachusetts Institute of Technology, McGovern Institute for Brain Research, Cambridge, MA 02139, USA.
Abstract:
We report several inductively coupled RF coil designs that are very easy to construct, produce high signal-to-noise ratio (SNR) and high spatial resolution while accommodating life support, anesthesia and monitoring in small animals. Inductively coupled surface coils were designed for hyperpolarized 13 C MR spectroscopic imaging (MRSI) of mouse brain, with emphases on the simplicity of the circuit design, ease of use, whole-brain coverage, and high SNR. The simplest form was a resonant loop designed to crown the mouse head for a snug fit to achieve full coverage of the brain with high sensitivity when inductively coupled to a broadband pick-up coil. Here, we demonstrated the coil's performance in hyperpolarized 13 C MRSI of a normal mouse and a glioblastoma mouse model at 4.7 T. High SNR exceeding 70:1 was obtained in the brain with good spatial resolution (1.53 mm × 1.53 mm). Similar inductively coupled loop for other X-nuclei can be made very easily in a few minutes and achieve high performance, as demonstrated in 31 P spectroscopy. Similar design concept was expanded to splitable, inductively coupled volume coils for high-resolution proton MRI of marmoset at 3T and 9.4T, to easily accommodate head restraint, vital-sign monitoring, and anesthesia delivery.
Related Concept Videos
Magnetic Resonance Imaging
¹H NMR Signal Multiplicity: Splitting Patterns
NMR Spectroscopy: Spin–Spin Coupling
Insensitive Nuclei Enhanced by Polarization Transfer (INEPT)
Atomic Nuclei: Magnetic Resonance
Applications Of NMR In Biology