Related Experiment Video
Updated: Jul 9, 2025

Radiation Planning Assistant - A Streamlined, Fully Automated Radiotherapy Treatment Planning System
Published on: April 11, 2018
Automating the Treatment Planning Process for Volumetric Modulated Arc Therapy Craniospinal Irradiation
Ignacio O Romero1, Eric A Simiele2, Yuliia Lozko1
1Department of Radiation Oncology, Stanford University, Stanford, California.
Purpose:
The purpose of this work is to develop a method to automate the treatment planning process of craniospinal irradiation (CSI) using volumetric modulated arc therapy.
Methods And Materials:
Two scripts were developed using the Eclipse Scripting Application Programming Interface to perform auto-plan preparation and optimization. Ten patients (age, 5-44 years) previously treated at our institution with low dose volumetric modulated arc therapy CSI (prescription of 12 Gy) before total body irradiation were selected to evaluate the efficacy of the proposed auto-planning process. Paired t tests compared the dosimetric indices of the auto-plans to the manually generated clinical plans. All plans were normalized to 95% of planning target volume (PTV) coverage with the prescription dose. Two physicians and one physicist were asked to evaluate the manual plans and auto-plans of each patient in a blinded retrospective review and to indicate clinical acceptability and which plans were preferred for treatment.
Results:
Compared with the manual CSI plans, the auto plans obtained significant reductions in Dmean to the parotids, submandibular glands, larynx, thyroid, and significant reduction in the plan PTV Dmax and D0.03 cc. The standard deviation range of the dosimetric parameters was greatly reduced for auto plans (range, 0.1-1.3 Gy) relative to manual plans (range, 0.4-5.9 Gy) indicating better plan consistency. Among the 10 patients, the auto-plans were preferred over the manual plans 90% of the time by the reviewing experts. The required time for auto-planning was approximately 1 hour compared with estimated 4 or more hours for manual planning.
Conclusions:
Reductions in planning time without sacrifices in plan quality were obtained using the auto-planning process compared with manual planning. Variation in plan quality was also reduced. The auto-planning scripts will be made freely available to other institutions and clinics.