A comparison of visual function tests in eyes with maculopathy.
Ophthalmology
|September 1, 1986
Summary
The Potential Acuity Meter (PAM) and focal electroretinogram (ERG) are most reliable for assessing visual function in patients with macular disease, outperforming other visual acuity tests. These findings aid in selecting appropriate diagnostic tools for maculopathy evaluation.
Area of Science:
- Ophthalmology
- Optometry
- Vision Science
Background:
- Accurate assessment of visual function is crucial for managing macular diseases.
- Several new tests aim to predict visual acuity in the presence of ocular conditions.
Purpose of the Study:
- To compare the diagnostic accuracy of recently developed visual function tests against Snellen acuity in eyes with macular disease.
- To determine the reliability of Potential Acuity Meter (PAM), laser interferometer (LI), white-light interferometer (WLI), blue field entoptic phenomenon, and focal electroretinogram (ERG) for evaluating macular function.
Main Methods:
- Compared visual acuity measurements from PAM, LI, WLI, blue field entoptic phenomenon, and focal ERG with Snellen acuity in 81 eyes with clear media and macular disease.
- Analyzed the agreement between test results and Snellen acuity, noting discrepancies in octaves and categorizing blue field and focal ERG as normal/abnormal.
Main Results:
- PAM, LI, and WLI tended to overestimate visual acuity compared to Snellen acuity.
- LI and WLI showed significant discrepancies (≥1.5 octaves) in ~40% of eyes.
- PAM demonstrated better agreement with Snellen acuity (91% within 1.5 octaves).
- Abnormal blue field and focal ERG results correlated with poor Snellen acuity (<20/40) in 65% and 91% of eyes, respectively.
Conclusions:
- PAM and focal ERG are the most reliable tests for evaluating macular function in the presence of maculopathy, assuming clear ocular media.
- These findings assist clinicians in selecting the most accurate visual function tests for patients with macular disease.