Predictors of the right ventricular perforation caused by active-fixation pacing and defibrillator leads: A

Umut Celikyurt1, Burak Acar1, Sadan Yavuz2

  • 1Arrhythmia, Electrophysiology, Pacemaker Research and Management Center, Department of Cardiology, Kocaeli University Medical Faculty, Umuttepe Yerleskesi, Kocaeli, Turkey.

Insights

Active-fixation leads can cause cardiac perforation. Radiographic predictors like lead slack and width during implantation can help identify patients at risk for right ventricular perforation.

Area of Science:

  • Cardiology
  • Medical Devices
  • Radiology

Background:

  • Active-fixation leads are linked to increased cardiac perforation rates.
  • Limited research exists on radiographic predictors of right ventricular (RV) lead perforation.

Purpose of the Study:

  • To identify radiographic predictors of RV lead perforation.
  • To evaluate the association between lead characteristics and perforation risk.

Main Methods:

  • Retrospective observational study of 1691 patients with active fixation pacing/defibrillator leads.
  • Analysis of 14 patients with clinically relevant RV perforation.

Main Results:

  • Lead slack score, change in lead slack height, and lead width were independent predictors of RV perforation in multivariate analysis.
  • Left ventricular ejection fraction also showed a significant association.

Conclusions:

  • Fluoroscopic predictors of RV perforation are identifiable during implantation.
  • Identifying these predictors may help prevent complications associated with active-fixation leads.
Abstract