Improved method robustness and ruggedness in liquid chromatography-mass spectrometry by increasing the acid content
A Kaufmann1, P Butcher1, K Maden1
1Official Food Control Authority of the Canton of Zurich, Kantonales Labor Zürich, Fehrenstrasse 15, 8032 Zürich, Switzerland.
Abstract:
A routine multiresidue method developed for the detection and quantification of veterinary drug residues in animal-based food was used to analyze sheep (ovine) liver. Unlike when working with previously validated matrices (e.g., bovine liver), some of the analytes of interest chromatographed in the form of split- or even fully baseline separated peaks. In other cases a significantly longer retention times (tR) was observed. A detailed investigation led to the elucidation of taurocholic acid as the causative agent. This compound is present in sheep liver at significantly higher concentrations than in most other animal tissues. Taurocholic acid is a zwitterionic compound and likely acts as an ion pairing agent, which modifies the selectivity of the stationary phase in a highly spatial and dynamic way. Injecting smaller volumes of matrix extract or the use of a significantly higher formic acid concentration in the mobile phase reduced or even completely eliminated the peak splitting. A more detailed examination led to the observation that the problem is not restricted to this particular matrix and extraction procedure or the used stationary phase. In fact, a higher formic acid concentration (e.g., 1.0 % versus 0.1 %) significantly improves the peak shape of many analytes present in fortified matrix samples as well as in pure standard solutions. In addition, analytical column aging was observed as being slower with a higher formic acid concentration. Finally the peak shape of analytes interacting with the metallic parts along the flow path of the liquid chromatograph was also significantly improved. Use of 0.1 % acid in mobile phases is often taken for granted in LC-MS. Regardless of the stationary phase, a higher ionic strength better stabilizes the pH and reduces unwanted interactions, which ultimately improves the method robustness. Flow injection experiments often show that 0.1 % acid concentrations produce the highest analyte signals. Yet, the use of 1 % acid in the mobile phase often leads to narrower and therefore taller chromatographic peaks, which may lead to lower detection limits for many analytes and to an improved separation efficiency.
More Related Videos
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
11:00Untargeted Metabolomics from Biological Sources Using Ultraperformance Liquid Chromatography-High Resolution Mass Spectrometry UPLC-HRMS
Published on: May 20, 2013
Related Concept Videos
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Instrumentation
Principles Of Column Chromatography
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...