Abstract:
Simulation of bioresorbable medical devices is hindered by the limitations of current material models. Useful simulations require that both the short- and long-term response must be considered; existing models are not physically-based and provide limited insight to guide performance improvements. This study presents an integrated degradation framework which couples a physically-based degradation model, which predicts changes in both crystallinity (Xc) and molecular weight (Mn), with the results of a micromechanical model, which predicts the effective properties of the semicrystalline polymer. This degradation framework is used to simulate the deployment of a bioresorbable PLLA (Poly (L-lactide) stent into a mock vessel and the subsequent mechanical response during degradation under different diffusion boundary conditions representing neointimal growth. A workflow is established in a commercial finite element code that couples both the immediate and long-term responses. Clinically relevant lumen loss is reported and used to compare different responses and the effect of neo-intimal tissue regrowth post-implantation on degradation and on the mechanical response is assessed. In addition, the effects of possible changes in Xc, which could occur during processing and stent deployment, are explored.