Image quality of photon-counting detector CT virtual monoenergetic and polyenergetic reconstructions for head and
Maria Vittoria Spampinato1, Jeffrey Rodgers2, Liam J McGill1
1Department of Radiology and Radiological Science, Medical University of South Carolina, 96 Jonathan Lucas Street, Charleston, SC 29425, United States of America.
Abstract:
We compared image quality of head and neck CT angiography (CTA) obtained with a photon-counting detector CT (PCD-CT), including virtual monoenergetic images and polyenergetic reconstructions, and conventional energy-integrating detectors CT (EID-CT) in three patients. PCD-CT monoenergetic reconstructions at 70 keV and lower provided excellent image quality, with improved signal-to-noise and contrast-to-noise compared to EID-CT and PCD-CT polyenergetic reconstructions. PCD-CT may enable radiation dose and iodinated contrast dose reduction for cerebrovascular imaging.