High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
Probe Electrospray Ionization Tandem Mass Spectrometry for the Detection and Quantification of Benzodiazepines
Pauline Griffeuille1, Souleiman El Balkhi1, Sylvain Dulaurent1
1Department of Pharmacology and Toxicology, Limoges University Hospital, Limoges ,France ; and.
Background:
Legally prescribed benzodiazepines (BZDs) and designer BZDs are widely misused and must be determined in multiple contexts (eg, overdose, drug-facilitated sexual assaults, or driving under the influence of drugs). This study aimed to develop a method for measuring serum BZD levels using probe electrospray ionization (PESI) mass spectrometry and an isotope dilution approach.
Methods:
A tandem mass spectrometer equipped with a probe electrospray ionization source in multiple reaction monitoring mode was used. Isotope dilution was applied for quantification using a deuterated internal standard at a fixed concentration for alprazolam, bromazepam, diazepam, nordiazepam, oxazepam, temazepam, zolpidem, and zopiclone. This method included designer BZDs: clonazolam, deschloroetizolam, diclazepam, etizolam, flualprazolam, flubromazepam, flubromazolam, meclonazepam, nifoxipam, and pyrazolam. Sample preparation was done by mixing 10 µL of serum with 500 µL of an ethanol/ammonium formate 0.01 mol/L buffer. Complete validation was performed, and the method was compared with liquid chromatography coupled with mass spectrometry (LC-MS/MS) and immunoassays (IC) by analyzing 40 real samples.
Results:
The analysis time for identification and quantification of the 18 molecules was 2.5 minutes. This method was fully validated, and the limits of quantification varied from 5 to 50 mcg/L depending on the molecule. In the 40 real samples, 100% of molecules (n = 89) were detected by both LC-MS/MS and PESI-MS/MS, and regression analysis showed excellent agreement between the 2 methods (r 2 = 0.98). On IC, bromazepam and zolpidem were not detected in 2 and 1 cases, respectively.
Conclusions:
PESI-MS/MS allows serum BZD detection and measurement. Given the isotope dilution approach, a calibration curve was not required, and its performance was similar to that of LC-MS/MS, and its specificity was higher than that of IC.
More Related Videos
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
High-Performance Liquid Chromatography: Types of Detectors
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Chemical Ionization (CI) Mass Spectrometry
Mass Analyzers: Common Types