Infrared Degenerate Four-wave Mixing with Upconversion Detection for Quantitative Gas Sensing
Published on: March 22, 2019
Mid-infrared trace detection with parts-per-quadrillion quantitation accuracy: Expanding frontiers of radiocarbon
1Center for Accelerator Mass Spectrometry, Atmospheric, Earth, and Energy Division, Physical and Life Sciences Directorate, Lawrence Livermore National Laboratory, Livermore, CA 94550.
Abstract:
Detection sensitivity is a critical characteristic to consider during selection of spectroscopic techniques. However, high sensitivity alone is insufficient for spectroscopic measurements in spectrally congested regions. Two-color cavity ringdown spectroscopy (2C-CRDS), based on intra-cavity pump-probe detection, simultaneously achieves high detection sensitivity and selectivity. This combination enables mid-infrared detection of radiocarbon dioxide ([Formula: see text]CO[Formula: see text]) molecules in room-temperature CO[Formula: see text] samples, with 1.4 parts-per-quadrillion (ppq, 10[Formula: see text]) sensitivity (average measurement precision) and 4.6-ppq quantitation accuracy (average calibrated measurement error for 21 samples from four separate trials) demonstrated on samples with [Formula: see text]C/C up to [Formula: see text]1.5[Formula: see text] natural abundance ([Formula: see text]1,800 ppq). These highly reproducible measurements, which are the most sensitive and quantitatively accurate in the mid-infrared, are accomplished despite the presence of orders-of-magnitude stronger, one-photon signals from other CO[Formula: see text] isotopologues. This is a major achievement in laser spectroscopy. A room-temperature-operated, compact, and low-cost 2C-CRDS sensor for [Formula: see text]CO[Formula: see text] benefits a wide range of scientific fields that utilize [Formula: see text]C for dating and isotope tracing, most notably atmospheric [Formula: see text]CO[Formula: see text] monitoring to track CO[Formula: see text] emissions from fossil fuels. The 2C-CRDS technique significantly enhances the general utility of high-resolution mid-infrared detection for analytical measurements and fundamental chemical dynamics studies.
More Related Videos
Related Concept Videos
IR Spectrometers
Attenuated Total Reflectance (ATR) Infrared Spectroscopy: Overview
The ATR process begins by directing a beam...
Infrared (IR) Spectroscopy: Overview
Different compounds display unique properties due to their...
Mass Analyzers: Common Types
¹³C NMR: ¹H–¹³C Decoupling
A broadband decoupling technique is used to simplify these complex, sometimes overlapping, signals. Broadband decoupling relies on a...
Gas Chromatography: Types of Detectors-II