A Precision Medicine Tool for Measurement and Monitoring of Hemoglobin S in Sickle Cell Disease Patients Receiving Transfusion Therapy
Technical Note: Comparison of Alinity c Hemoglobin A1c Immunoassay with Premier Hb9210 Automated HPLC Assay: A
Ashton Walton1, Jonathan Meador1, Kelsey Woodard1
1Clinical Laboratories, The University of Kansas Hospital, Kansas City, KS, USA.
Objective:
We utilized Premier Hb9210 analyzer (HPLC method; Trinity Biotech, Jamestown, NY) for measuring HBA1c in whole blood. As our laboratory is transitioning to Abbott system, we compared HbA1c values obtained by Alinity c and Premier Hb9210.
Methods:
The Premier Hb9210 analyzer is based on boronate affinity high performance liquid chromatography with analytical measurement range of 3.8 to 18.5%. The Alinity c Hemoglobin A1c assay measured both total hemoglobin and HbA1c (enzymatic assay) in whole blood and then calculated %HbA1c. The analytical measurement range of this assay is 4 to 14% of HbA1c. We evaluated the analytical performance of Alinity c HbA1c by evaluating precision and also comparing 77 clinical samples with our reference HPLC method.
Results:
Both Alinity c HbA1c and Premier HB9210 have excellent total precision. Plotting HbA1c results obtained by the Premier Hb9210 analyzer in the x-axis (currently used reference method) and the corresponding values obtained by the Alinity c assay, we observed the following regression equation: y=0.9473x+0.1548 ( n=77, r=0.99).
Discussion:
Our result indicates that HbA1c enzymatic assay on the Alinity c analyzer showed values comparable to HPLC method. However, at the decision points (2.8% average negative bias at >6.4% and 3.3% average negative bias at 7%), HbA1c values obtained by the Alinity c analyzer were lower than the reference method.
Conclusions:
We conclude that HbA1c assay on the Alinity c analyzer is a viable alternative to HPLC for measuring HbA1c in clinical laboratories but values at the decision points must be interpreted with caution and if necessary should be repeated by a reference HPLC method.