Implementation of a Reference Interferometer for Nanodetection
Published on: April 26, 2014
Self-Mixing Interferometer for Acoustic Measurements through Vibrometric Calibration
Simon Chanu-Rigaldies1, Pierre Lecomte2, Sébastien Ollivier2
1Ecole Centrale de Lyon, CNRS, Universite Claude Bernard Lyon 1, INSA Lyon, LMFA, UMR5509, 69130 Ecully, France.
This study demonstrates using a Self-Mixing Interferometer (SMI) for accurate Sound Pressure Level (SPL) measurements in acoustic waveguides. The calibrated SMI sensor achieved results comparable to reference microphones across a wide frequency and amplitude range.
Area of Science:
- Acoustics
- Optical Engineering
- Metrology
Background:
- Self-Mixing Interferometers (SMI) are self-aligned optical devices.
- SMIs have been utilized for acoustic wave sensing via the acousto-optic effect.
Purpose of the Study:
- To adapt and validate the use of an SMI for measuring Sound Pressure Level (SPL) within acoustic waveguides.
- To establish a calibration method for SMI sensors as acoustic pressure sensors.
Main Methods:
- In situ calibration of the SMI as a vibrometer.
- Estimation of optical feedback parameters (C and α) by solving nonlinear SMI behavior equations.
- Validation using simulated SMI signals and comparison with a reference microphone.
Main Results:
- Accurate estimation of SMI optical feedback parameters in the strong feedback regime (C≥4.6).
- SPL measurements using the calibrated SMI showed a maximal deviation of 2.2 dB compared to a reference microphone.
- Successful SPL measurement across a broad range of plane wave amplitudes (20–860 Pa) and frequencies (614–17,900 Hz).
Conclusions:
- The Self-Mixing Interferometer can be effectively calibrated and utilized as a precise acoustic pressure sensor.
- The method provides reliable SPL measurements in acoustic waveguides, demonstrating its practical applicability.
- The study validates the SMI's potential for advanced acoustic sensing applications.
Related Concept Videos
IR Spectrum Peak Splitting: Symmetric vs Asymmetric Vibrations
Electronic Distance Measuring Instruments
Interference and Diffraction
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...