Headpulse measurement can reliably identify large-vessel occlusion stroke in prehospital suspected stroke patients:
James H Paxton1, Kevin J Keenan2, John M Wilburn1
1Department of Emergency Medicine, Wayne State University School of Medicine, Detroit, Michigan, USA.
Background:
Large-vessel occlusion (LVO) stroke represents one-third of acute ischemic stroke (AIS) in the United States but causes two-thirds of poststroke dependence and >90% of poststroke mortality. Prehospital LVO stroke detection permits efficient emergency medical systems (EMS) transport to an endovascular thrombectomy (EVT)-capable center. Our primary objective was to determine the feasibility of using a cranial accelerometry (CA) headset device for prehospital LVO stroke detection. Our secondary objective was development of an algorithm capable of distinguishing LVO stroke from other conditions.
Methods:
We prospectively enrolled consecutive adult patients suspected of acute stroke from 11 study hospitals in four different U.S. geographical regions over a 21-month period. Patients received device placement by prehospital EMS personnel. Headset data were matched with clinical data following informed consent. LVO stroke diagnosis was determined by medical chart review. The device was trained using device data and Los Angeles Motor Scale (LAMS) examination components. A binary threshold was selected for comparison of device performance to LAMS scores.
Results:
A total of 594 subjects were enrolled, including 183 subjects who received the second-generation device. Usable data were captured in 158 patients (86.3%). Study subjects were 53% female and 56% Black/African American, with median age 69 years. Twenty-six (16.4%) patients had LVO and 132 (83.6%) were not LVO (not-LVO AIS, 33; intracerebral hemorrhage, nine; stroke mimics, 90). COVID-19 testing and positivity rates (10.6%) were not different between groups. We found a sensitivity of 38.5% and specificity of 82.7% for LAMS ≥ 4 in detecting LVO stroke versus a sensitivity of 84.6% (p < 0.0015 for superiority) and specificity of 82.6% (p = 0.81 for superiority) for the device algorithm (CA + LAMS).
Conclusions:
Obtaining adequate recordings with a CA headset is highly feasible in the prehospital environment. Use of the device algorithm incorporating both CA and LAMS data for LVO detection resulted in significantly higher sensitivity without reduced specificity when compared to the use of LAMS alone.
More Related Videos
Related Concept Videos
Pulse Assessment Sites
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring pulse
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pulse
The pulse serves as a clinical...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: