Spatial and temporal motion characterization for x-ray CT
1Independent Consultant, Brookfield, Wisconsin, USA.
Background:
Motion induced image artifacts have been the focus of many investigations for x-ray computed tomography (CT). Methodologies of combating patient motion include the use of gating devices to optimize the data acquisition, reduction in patient scan time via faster gantry rotation and large detector coverage, and the development of advanced reconstruction and post-processing algorithms to minimize motion artifacts.
Purpose:
Previously proposed approaches are generally "global" in nature in that motion is characterized for the entire image. It is well known, however, that the presence of motion artifact in a CT image is highly nonuniform. When there is a lack of automated and quantitative local measure indicating the presence and the severity of motion artifacts in a local region, the quality of the reconstructed images depends heavily on the CT operator's rigor and experience. Even when an operator is informed of the presence of motion, little information is provided about the nature of the motion artifact to understand its relevance to the clinical task at hand. In this paper, we propose an image-space spatial- and temporal-consistency metric (CM) to detect and characterize the local motion.
Method:
In a non-rigid human organ, such as the lung, there are many small and rigid objects (target objects), such as blood vessels and nodules, distributed throughout the organ. If motion can be characterized for these target objects, we obtain a complete motion map for the organ. To accomplish this, a preliminary image reconstruction is carried out to identify the target objects and establish region-of-interests for consistency-metric calculation. The CM is then obtained based on the backprojected intensity difference between the object region and its circular background. For a stationary object, the accumulation of this quantity over views is linear. When a target object moves, nonlinear behavior exhibits and a quantitative measure of linearity indicates the severity of motion.
Results:
Extensive computer simulation was utilized to confirm the validity of the theory. These tests stress the sensitivity of the proposed CM to the target object size, object shape, in-plane motion, cross-plane motion, cone-beam effect, and complex background. Results confirm that the proposed approach is robust under different testing conditions. The proposed CM is further validated using a cardiac scan of a swine, and the proposed CM correlates well with the visual inspection of the artifact in the reconstructed images.
Conclusions:
In this paper, we have demonstrated the efficacy of the proposed CM for motion detection. Unlike previously proposed approaches where the consistency condition is derived for the entire image or the entire imaging volume, the proposed metric is well localized so that different zones in a patient anatomy can be individually characterized. In addition, the proposed CM provides a quantitative measure on a view-by-view basis so that the severity of motion is consistently estimated over time. Such information can be used to optimize the image reconstruction process and minimize the motion artifact.
More Related Videos
09:00Visualization of Failure and the Associated Grain-Scale Mechanical Behavior of Granular Soils under Shear using Synchrotron X-Ray Micro-Tomography
Published on: September 29, 2019
10:18Dynamic Pore-scale Reservoir-condition Imaging of Reaction in Carbonates Using Synchrotron Fast Tomography
Published on: February 21, 2017
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
X-ray Imaging
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...