Deep silicon photon-counting CT: A first simulation-based study for assessing perceptual benefits across diverse
Shobhit Sharma1, Debashish Pal2, Ehsan Abadi3
1Center for Virtual Imaging Trials and Carl E. Ravin Advanced Imaging Laboratories, 2424 Erwin Rd, Suite 302, Durham, NC 27705, USA; Department of Physics, Duke University, Science Drive, Durham, NC 27708, USA.
Objectives:
To assess perceptual benefits provided by the improved spatial resolution and noise performance of deep silicon photon-counting CT (Si-PCCT) over conventional energy-integrating CT (ECT) using polychromatic images for various clinical tasks and anatomical regions.
Materials And Methods:
Anthropomorphic, computational models were developed for lungs, liver, inner ear, and head-and-neck (H&N) anatomies. These regions included specific abnormalities such as lesions in the lungs and liver, and calcified plaques in the carotid arteries. The anatomical models were imaged using a scanner-specific CT simulation platform (DukeSim) modeling a Si-PCCT prototype and a conventional ECT system at matched dose levels. The simulated polychromatic projections were reconstructed with matched in-plane resolutions using manufacturer-specific software. The reconstructed pairs of images were scored by radiologists to gauge the task-specific perceptual benefits provided by Si-PCCT compared to ECT based on visualization of anatomical and image quality features. The scores were standardized as z-scores for minimizing inter-observer variability and compared between the systems for evidence of statistically significant improvement (one-sided Wilcoxon rank-sum test with a significance level of 0.05) in perceptual performance for Si-PCCT.
Results:
Si-PCCT offered favorable image quality and improved visualization capabilities, leading to mean improvements in task-specific perceptual performance over ECT for most tasks. The improvements for Si-PCCT were statistically significant for the visualization of lung lesion (0.08 ± 0.89 vs. 0.90 ± 0.48), liver lesion (-0.64 ± 0.37 vs. 0.95 ± 0.55), and soft tissue structures (-0.47 ± 0.90 vs. 0.33 ± 1.24) and cochlea (-0.47 ± 0.80 vs. 0.38 ± 0.62) in inner ear.
Conclusions:
Si-PCCT exhibited mean improvements in task-specific perceptual performance over ECT for most clinical tasks considered in this study, with statistically significant improvement for 6/20 tasks. The perceptual performance of Si-PCCT is expected to improve further with availability of spectral information and reconstruction kernels optimized for high resolution provided by smaller pixel size of Si-PCCT. The outcomes of this study indicate the positive potential of Si-PCCT for benefiting routine clinical practice through improved image quality and visualization capabilities.