Related Experiment Video
Updated: Jun 27, 2025

An Intramedullary Locking Nail for Standardized Fixation of Femur Osteotomies to Analyze Normal and Defective Bone Healing in Mice
Published on: November 13, 2016
Finite element method analysis of bone stress for variants of locking plate placement
Marek Palka1, Patrycja Miszczyk2, Maciej Jurewicz3
1Institute of Aeronautics and Applied Mechanics, Warsaw University of Technology, Nowowiejska 24, 00-660, Warsaw, Poland.
Abstract:
This study investigates the optimal placement of locking plate screws for bone fracture stabilization in the humerus, a crucial factor for enhancing healing outcomes and patient comfort. Utilizing Finite Element Method (FEM) modeling, the research aimed to determine the most effective screw configuration for achieving optimal stress distribution in the humerus bone. A computer tomography (CT) scan of the humerus was performed, and the resulting images were used to create a detailed model in SOLIDWORKS 2012. This model was then analyzed using ANSYS Workbench V13 to develop a finite element model of the bone. Four different screw configurations were examined: 4 × 0°, 4 × 10°, 4 × 20°, 2 × 20°; 2 × 0°. These configurations were subjected to bending in the XZ and YZ planes, as well as tension and compression along the Z axis. The research identified the 2 × 20°+2 × 0° configuration as the most beneficial, with average stress values below 30 MPa and peak stress values below 50 MPa in 3-point bending at the first screw. This configuration consistently showed the lowest stress values across all loading scenarios. Specifically, stress levels ranged from 20 MPa to 50 MPa for bending in the XZ plane, 20 MPa-35 MPa for bending in the YZ plane, 20 MPa-30 MPa for extension in the Z-axis, and 18 MPa-25 MPa for compression in the Z-axis. The 4 × 10° and 4 × 20° configurations also produced satisfactory results, with stress levels not exceeding 70 MPa. However, the 4 × 0° configuration presented considerable stress during bending and compression in the Z-axis, with stress values exceeding 100 MPa, potentially leading to mechanical damage. In conclusion, the 2 × 20°; 2 × 0° screw configuration was identified as the most effective in minimizing stress on the treated bone. Future work will involve a more detailed analysis of this methodology and its potential integration into clinical practice, with a focus on enhancing patient outcomes in bone fracture treatment.
More Related Videos
09:32Subject-specific Musculoskeletal Model for Studying Bone Strain During Dynamic Motion
Published on: April 11, 2018
15:11Surgical Fixation of Sternal Fractures: Preoperative Planning and a Safe Surgical Technique Using Locked Titanium Plates and Depth Limited Drilling
Published on: January 5, 2015
Related Concept Videos
Saint-Venant's Principle
Plastic Deformations of Members with a Single Plane of Symmetry
Flexural Stress
Hooke's Law states that within the material's elastic limits, stress is directly proportional to strain. In a member experiencing a bending moment, the strain at any point is relative to...
Stresses under Combined Loadings
The process begins by slicing the tube at critical points and analyzing the internal forces and stress components at these sections, focusing on the centroid. Normal stresses, generated by axial forces and bending moments, are either compressive or tensile and vary across the section from...
Temperature Dependent Deformation
Relation between Poisson's ratio, Modulus of Elasticity and Modulus of Rigidity