A channel-homogenized micro gas chromatography column for light alkane separation with uniform HKUST-1 stationary
Shaojie Ma1, Yuchen Zhu1, Boxin Chen1
1State Key Laboratory of Transducer Technology, Shanghai Institute of Microsystem and Information Technology, Chinese Academy of Sciences, Shanghai 200050, China; Center of Materials Science and Optoelectronics Engineering, University of Chinese Academy of Sciences, Beijing 100049, China.
Background:
The micro gas chromatography column (μGCC) is one of the key components of the miniaturized gas chromatography system. However, light alkanes are difficult to be separated by a micro gas chromatography column, especially for methane and ethane, because the length of μGCC is limited by the area of a silicon substrate. More importantly, the heterogeneous microchannel surface formed by silicon glass bonding causes uneven stationary phase coating and the forces between the untreated microchannel surfaces and the stationary phase materials are weak, which will prevent the improvement of separation performance.
Results:
In this paper, a micro gas chromatography column (μGCC) with uniform HKUST-1 stationary phase is reported. Significantly, an alumina film prepared by the atomic layer deposition (ALD) technique is used to homogenize the heterogeneous microchannels. The alumina is a hydrophilic material and the alumina made by the ALD technique is uniform. The forces between hydrophilic alumina film and HKUST-1 are strong, which can greatly improve the coating uniformity of the hydrophilic stationary phase HKUST-1. The test results show that the μGCC could baseline separate the light alkane mixtures (CH4, C2H6, C3H8, and C4H10) at the high testing temperature of 120 °C. The maximum resolution of the difficult-separated methane and ethane reached 19.2, which is 108 % higher than the μGCC using the same stationary phase without homogenizing the microchannel inner surface.
Significance:
The μGCC uses ALD alumina film to homogenize the microchannel inner surface; meanwhile, hydrophilic ALD alumina has a strong electrostatic attraction with the hydrophilic stationary phase HKUST-1. Homogeneous microchannel surface and strong electrostatic attraction are favorable to obtain uniform stationary phase which greatly improves the separation performance, resulting in a large resolution for methane and ethane. The μGCC has broad application prospects in light alkane separation.
More Related Videos
10:14Chromatographic Fingerprinting by Template Matching for Data Collected by Comprehensive Two-Dimensional Gas Chromatography
Published on: September 2, 2020
08:37Measurement of H2S in Crude Oil and Crude Oil Headspace Using Multidimensional Gas Chromatography, Deans Switching and Sulfur-selective Detection
Published on: December 10, 2015
Related Concept Videos
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Instrumentation