Magnetic Resonance Elastography Methodology for the Evaluation of Tissue Engineered Construct Growth
Published on: February 9, 2012
Free-breathing motion corrected magnetic resonance elastography of the abdomen
Anne-Sophie van Schelt1,2, Nienke Petronella Maria Wassenaar1,2, Jurgen Henk Runge1,3
1Department of Radiology and Nuclear Medicine, Amsterdam UMC, University of Amsterdam, Amsterdam, The Netherlands.
Background:
Magnetic resonance elastography (MRE) is a non-invasive method to measure the viscoelastic properties of tissue and has been applied in multiple abdominal organs. However, abdominal MRE suffers from detrimental breathing motion causing misalignment of structures between repeated acquisitions for different MRE dimensions (e.g., motion encoding directions and wave phase offsets). This study investigated motion correction strategies to resolve all breathing motion on sagittal free-breathing MRE acquisitions in a phantom, in healthy volunteers and showed feasibility in patients.
Methods:
First, in silico experiments were performed on a static phantom dataset with simulated motion. Second, eight healthy volunteers underwent two sagittal MRE acquisitions in the pancreas and right kidney. The multi-frequency free-breathing spin-echo echo-planar-imaging (SE-EPI) MRE consisted of four frequencies (30, 40, 50, 60 Hz), eight wave-phase offsets, with 3 mm3 isotropic voxel size. Following data re-sorting in different number of motion states (4 till 12) based on respiratory waveform signal, three intensity-based registration methods (monomodal, multimodal, and phase correlation) and non-rigid local registration were compared. A ranking method was used to determine the best registration method, based on seven signal-to-noise and image quality measures. Repeatability was assessed for no motion correction (Original) and the best performing method (Best) using Bland-Altman analysis. Lastly, the best motion correction method was compared to no motion correction on patient MRE data [pancreatic ductal adenocarcinoma (PDAC, n=5) and metabolic dysfunction-associated steatotic liver disease (MASLD) (n=1)].
Results:
In silico experiments showed a deviation of shear wave speed (SWS) with simulated motion to the ground truth, which was (partially) resolved using motion correction. In healthy volunteers ranking resulted in the best motion correction method of monomodal registration using nine motion states, while no motion correction was ranked last. Limits of agreement were (-0.18, 0.14), and (-0.25, 0.18) m/s for Best and Original, respectively. Using motion correction in patients resulted in a significant increase in SWS in the pancreas (Original: 1.39±0.10 and Best: 1.50±0.17 m/s). After motion correction PDAC had a mean SWS of 1.56±0.27 m/s (Original: 1.42±0.25 m/s). The fibrotic liver mean SWS was 2.07±0.20 m/s (Original: 2.12±0.18 m/s).
Conclusions:
Motion correction in sagittal free-breathing abdominal MRE results in improved data quality, inversion precision, repeatability, and is feasible in patients.
Related Concept Videos
Magnetic Resonance Imaging
Assessment of the Abdomen I: Inspection and Auscultation
The abdominal examination is a cornerstone of clinical medicine, serving as a critical tool in diagnosing various gastrointestinal (GI) diseases. It involves a systematic approach that includes inspection and auscultation, each with distinct yet complementary roles in assessing the abdomen. This article will delve into these two primary methods healthcare professionals use to examine the abdomen.
Inspection of the Abdomen
The first step in any abdominal examination is inspection....
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Assessment of the Abdomen II: Percussion
Percussion
Percussion is an essential...