Evaluation of a new transpalpebral tonometer for self-measuring intraocular pressure
1Toneasy Inc., Tokyo, Japan.
Purpose:
We developed a novel transpalpebral self-tonometer called the TapEye tonometer (TET) based on palpation of the upper eyelid. Our goal was to evaluate a method for improving the accuracy of measuring intraocular pressure (IOP) through the eyelid.
Methods:
Participants underwent standardized training by technicians and were required to be able to use the TET for study inclusion. Subsequently, a noncontact tonometer and Goldmann applanation tonometer (GAT) were used. All participants were instructed to measure their IOPs using the three tonometers at baseline (visit 1) and at 1 month (visit 2). At visit 2, the corrected IOP value measured by the TET (c-TET) was calculated using the difference between the TET and GAT measurements obtained at visit 1.
Results:
No significant correlations were found between the TET and GAT measurements at any visit, but the correlation between the c-TET and GAT measurements at visit 2 was significant. The mean difference between the c-TET and GAT measurements was 0.4 ± 3.7 mmHg in the right eye and 0.5 ± 3.4 mmHg in the left eye.
Conclusions:
After correcting the IOP based on the difference between the TET and GAT measurements at the initial visit, the corrected IOP value of the TET was correlated with that of the GAT at the second visit. The TET has the potential to address an unmet need by providing a tool for minimally invasive IOP measurements.
Trial Registration:
Clinical trial registration number: jRCTs032220268.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Measurement of Blood Pressure