Design and Biocompatibility of a Novel, Flexible Artificial Cornea
Gavin Li1,2, Anthony J Aldave3, Guillermo Amescua4
1The Ocular Surface Disease Clinic, The Wilmer Eye Institute, Johns Hopkins University School of Medicine, Baltimore, MD, USA.
Purpose:
We sought to introduce the materials, design, and biocompatibility of a flexible and suturable artificial corneal device.
Methods:
Single-piece, fully synthetic, optic-skirt design devices were made from compact perfluoroalkoxy alkane. The skirt and the optic wall surfaces were lined with a porous tissue ingrowth material using expanded polytetrafluoroethylene. Full-thickness macroapertures around the skirt perimeter were placed to facilitate nutrition of the recipient cornea. Material properties including the skirt's modulus of elasticity and bending stiffness, optic light transmission, wetting behavior, topical drug penetrance, and degradation profile were evaluated.
Results:
The final prototype suitable for human use has a transparent optic with a diameter of 4.60 mm anteriorly, 4.28 mm posteriorly, and a skirt outer diameter of 6.8 mm. The biomechanical and optical properties of the device closely align with the native human cornea with an average normalized device skirt-bending stiffness of 4.7 kPa·mm4 and light transmission in the visible spectrum ranging between 92% and 96%. No optical damage was seen in the 36 devices tested in fouling experiments. No significant difference was observed in topical drug penetrance into the anterior chamber of the device implanted eye compared with the naïve rabbit eye.
Conclusions:
The flexibility and biocompatibility of our artificial cornea device may offer enhanced tissue integration and decreased inflammation, leading to improved retention compared with rigid keratoprosthesis designs.
Translational Relevance:
We have developed a fully synthetic, flexible, suturable, optic-skirt design prototype artificial cornea that is ready to be tested in early human feasibility studies.