Effects and avoidance of photoconversion-induced artifacts in confocal and STED microscopy
Anindita Dasgupta1,2, Agnes Koerfer1,2, Boštjan Kokot3
1Institute for Applied Optics and Biophysics, Friedrich Schiller University Jena, Jena, Germany.
None:
Fluorescence microscopy is limited by photoconversion due to continuous illumination, which results in not only photobleaching but also conversion of fluorescent molecules into species of different spectral properties through photoblueing. Here, we determined different fluorescence parameters of photoconverted products for various fluorophores under standard confocal and stimulated emission depletion (STED) microscopy conditions. We observed changes in both fluorescence spectra and lifetimes that can cause artifacts in quantitative measurements, which can be avoided by using exchangeable dyes.
Related Concept Videos
Confocal Fluorescence Microscopy
Super-resolution Fluorescence Microscopy
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...