Related Experiment Video
Updated: Jun 24, 2025

Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
Automation framework for online review of daily dose distributions using CT on Rails in proton therapy
Christian Dial1, Geoff Nelson1, Vikren Sarkar1
1Department of Radiation Oncology, University of Utah, 1950 Circle of Hope, Salt Lake City, UT, 84112, USA.
Purpose:
To demonstrate a framework for calculating daily dose distributions for proton therapy in a timeframe amenable to online evaluation using CT-on-Rails.
Methods:
Tasks associated with calculation of daily dose are fully automated. A rigid registration between daily and planning images is used to propagate beams and targets for calculation of daily dose; additionally, risk structures are propagated using deformable registration to facilitate online evaluation. An end-to-end constancy test was carried out using a pelvis phantom containing a simulated target and bladder contour. 97 Daily fan-beam CT data sets associated with 10 clinical patients were processed to demonstrate feasibility and utility of online evaluation. Computing times and dosimetric differences are reported.
Results:
The phantom constancy test took 62 s to complete with no notable discrepancies in the registrations or calculated dose. Max doses were identical for target and bladder contours on initial and repeat scans (359 and 310 cGy (RBE) respectively). Total processing time for 97 daily patient images averaged 154.6 s (73.0 - 222.0 s; SD = 31.8 s). On average, dose calculation accounted for 35 % of total processing time. Average differences in D95 for target contours was 1.5 % (SD = 1.6 %) with a max decrease of 5.9 % on a single daily image.
Conclusion:
Daily dose can be automatically calculated in a timeframe amenable to online evaluation using scanner utilities in conjunction with the scripting API of a commercial treatment planning system. Online evaluation of dose in proton therapy is useful to detect clinically relevant changes, guide setup, and facilitate treatment or replanning decisions.
More Related Videos
05:18Author Spotlight: Improving Radiation Therapy Access with Radiation Planning Assistant
Published on: October 6, 2023
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...