New Device for the Correction of Aphakia Without Capsular Support.
Victor Dias Bergamasco1, Vagner Rogério Dos Santos1, Bernardo Kaplan Moscovici1,2
1Department of Ophthalmology and Visual Sciences, Federal University of São Paulo (UNIFESP), São Paulo, Brazil.
Current Eye Research
|June 10, 2024
Summary
A new 3D-printed prototype for aphakia without capsular support (AWCS) was developed and tested. While the device showed promise, achieving optimal intraocular lens (IOL) stability and centralization remains a challenge.
Area of Science:
- Ophthalmology
- Biomedical Engineering
- Medical Device Development
Background:
- Aphakia without capsular support (AWCS) presents a surgical challenge for intraocular lens (IOL) implantation.
- Existing methods for managing AWCS may have limitations in achieving stable IOL positioning.
- Novel approaches are needed to address the complexities of IOL fixation in the absence of a capsular bag.
Purpose of the Study:
- To describe a novel 3D-printed prototype device for managing aphakia without capsular support (AWCS).
- To demonstrate the proof of concept for this new AWCS management technology.
- To evaluate the feasibility and initial performance of the prototype device and its implantation technique.
Main Methods:
- Virtual prototyping using 3D software for design and analysis.
- Manufacturing of the prototype using a nylon-6/nylon-6.6 copolymer filament and a 3D printer.
- Development of a surgical implantation technique utilizing a 23-gauge hypodermic needle and specific scleral marking guidelines.
- Haptic modification involving cutting and cauterization for scleral anchoring and flange formation.
Main Results:
- Successful prototyping and development of an implantation technique for AWCS.
- No vitreous tissue extrusion observed during sclerotomy, indicating a safe initial procedure.
- The device demonstrated good initial fixation to the sclera.
- Challenges remain in achieving adequate IOL stability and centralization post-implantation.
Conclusions:
- The development of a 3D-printed technology prototype for correcting aphakia without capsular support is feasible.
- The prototype addresses several key design considerations for AWCS management.
- Further refinement is necessary to overcome challenges related to IOL stability and centralization for optimal clinical outcomes.