Intan Technologies integrated circuits can produce analog-to-digital conversion artifacts that affect neural signal
Katrina Barth1, Cecilia Schmitz1, Thomas Jochum1
1Department of Biomedical Engineering, Duke University, Durham, NC, United States of America.
None:
Objective.Intan Technologies' integrated circuits (ICs) are valuable tools for neurophysiological data acquisition, providing signal amplification, filtering, and digitization from many channels (up to 64 channels/chip) at high sampling rates (up to 30 kSPS) within a compact package (⩽9× 7 mm). However, we found that the analog-to-digital converters (ADCs) in the Intan RHD2000 series ICs can produce artifacts in recorded signals. Here, we examine the effects of these ADC artifacts on neural signal quality and describe a method to detect them in recorded data.Approach.We identified two types of ADC artifacts produced by Intan ICs: 1) jumps, resulting from missing output codes, and 2) flatlines, resulting from overrepresented output codes. We identified ADC artifacts in neural recordings acquired with Intan RHD2000 ICs and tested the repeated performance of 17 ICsin vitro. With the on-chip digital-signal-processing disabled, we detected the ADC artifacts in each test recording by examining the distribution of unfiltered ADC output codes.Main Results.We found larger ADC artifacts in recordings using the Intan RHX data acquisition software versions 3.0-3.2, which did not run the necessary ADC calibration command when the inputs to the Intan recording controller were rescanned. This has been corrected in the Intan RHX software version 3.3. We found that the ADC calibration routine significantly reduced, but did not fully eliminate, the occurrence and size of ADC artifacts as compared with recordings acquired when the calibration routine was not run (p< 0.0001). When the ADC calibration routine was run, we found that the artifacts produced by each ADC were consistent over time, enabling us to sort ICs by performance.Significance.Our findings call attention to the importance of evaluating signal quality when acquiring electrophysiological data using Intan Technologies ICs and offer a method for detecting ADC artifacts in recorded data.
More Related Videos
10:58A Procedure for Implanting Organized Arrays of Microwires for Single-unit Recordings in Awake, Behaving Animals
Published on: February 14, 2014
09:00Investigating the Function of Deep Cortical and Subcortical Structures Using Stereotactic Electroencephalography: Lessons from the Anterior Cingulate Cortex
Published on: April 15, 2015
Related Concept Videos
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
Reconstruction of Signal using Interpolation