Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Pipe Flowrate Measurement01:28

Pipe Flowrate Measurement

662
In pipe flow measurement, orifice, nozzle, and Venturi meters are commonly used to determine fluid flowrates by constricting the flow area, which increases fluid velocity and reduces pressure. This pressure difference, governed by Bernoulli's principle and adjusted for real-world conditions, is essential for calculating flowrate. Each meter type is suited to specific applications based on accuracy, efficiency, and compatibility with various flow conditions.
The orifice meter is a simple,...
662
Assessment of Ventilation I: Respiratory Rate01:20

Assessment of Ventilation I: Respiratory Rate

1.1K
Assessment of Ventilation
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
1.1K
Equipments Used To Measure Blood Pressure01:30

Equipments Used To Measure Blood Pressure

957
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
957
Constant Volume Calorimetry02:41

Constant Volume Calorimetry

27.1K
Calorimeters are useful to determine the heat released or absorbed by a chemical reaction. Coffee cup calorimeters are designed to operate at constant (atmospheric) pressure and are convenient to measure heat flow (or enthalpy change) accompanying processes that occur in solution at constant pressure. A different type of calorimeter that operates at constant volume, colloquially known as a bomb calorimeter, is used to measure the energy produced by reactions that yield large amounts of heat and...
27.1K

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

McArdle disease (Glycogen Storage Disease Type V) from an exercise physiology and biochemistry perspective: Important considerations for exercise and physical activity testing and data interpretation.

Sports medicine and health science·2026
Same author

D- and L-lactate dehydrogenase reactions increase pH to lactate production and decrease pH to pyruvate production.

Frontiers in molecular biosciences·2026
Same author

<math><mrow><mover><mi>V</mi> <mo>˙</mo></mover></mrow></math> O<sub>2</sub> linear-onset kinetics spanning steady- and non-steady-state exercise.

Frontiers in physiology·2025
Same author

The missing hydrogen ion, Part-3: Science and the human flaws that compromise it.

Sports medicine and health science·2024
Same author

The missing hydrogen ion, part-2: Where the evidence leads to.

Sports medicine and health science·2024
Same author

The missing hydrogen ion, part-1: Historical precedents vs. fundamental concepts.

Sports medicine and health science·2024

Related Experiment Video

Updated: Jun 23, 2025

Calorespirometry: A Powerful, Noninvasive Approach to Investigate Cellular Energy Metabolism
09:08

Calorespirometry: A Powerful, Noninvasive Approach to Investigate Cellular Energy Metabolism

Published on: May 31, 2018

7.5K

Evaluating Airflow Sensor Methods: Precision in Indirect Calorimetry.

Praneel Titheradge1, Robert Robergs2

  • 1School of Allied Health, Exercise & Sports Sciences, Charles Sturt University, Port Macquarie, New South Wales, Australia.

Scandinavian Journal of Medicine & Science in Sports
|June 19, 2024
PubMed
Summary
This summary is machine-generated.

Different airflow sensors impact indirect calorimetry accuracy during exercise. Exercise intensity affects measurements, but sensor type does not significantly alter results, though agreement decreases at higher ventilation rates.

Keywords:
heterogeneityindirect calorimetryresponderstechnical error

More Related Videos

Combining Volumetric Capnography And Barometric Plethysmography To Measure The Lung Structure-function Relationship
08:25

Combining Volumetric Capnography And Barometric Plethysmography To Measure The Lung Structure-function Relationship

Published on: January 8, 2019

9.4K
Author Spotlight: Integrating Alveolar-Capillary Reserve Measurements in Exercise Adaptation and Therapeutic Strategies
08:44

Author Spotlight: Integrating Alveolar-Capillary Reserve Measurements in Exercise Adaptation and Therapeutic Strategies

Published on: February 2, 2024

696

Related Experiment Videos

Last Updated: Jun 23, 2025

Calorespirometry: A Powerful, Noninvasive Approach to Investigate Cellular Energy Metabolism
09:08

Calorespirometry: A Powerful, Noninvasive Approach to Investigate Cellular Energy Metabolism

Published on: May 31, 2018

7.5K
Combining Volumetric Capnography And Barometric Plethysmography To Measure The Lung Structure-function Relationship
08:25

Combining Volumetric Capnography And Barometric Plethysmography To Measure The Lung Structure-function Relationship

Published on: January 8, 2019

9.4K
Author Spotlight: Integrating Alveolar-Capillary Reserve Measurements in Exercise Adaptation and Therapeutic Strategies
08:44

Author Spotlight: Integrating Alveolar-Capillary Reserve Measurements in Exercise Adaptation and Therapeutic Strategies

Published on: February 2, 2024

696

Area of Science:

  • Exercise Physiology
  • Metabolic Measurement
  • Biomedical Engineering

Background:

  • Indirect calorimetry is crucial for assessing energy expenditure during exercise.
  • Accurate measurement of pulmonary ventilation is essential for reliable indirect calorimetry.
  • Various airflow sensors exist, each with potential impacts on measurement precision.

Purpose of the Study:

  • To evaluate the influence of three distinct volumetric gas flow measurement methods (turbine, pneumotachograph, Venturi) on the predictive accuracy and precision of expired gas analysis indirect calorimetry (EGAIC).
  • To determine how varying exercise intensities affect the performance of these airflow measurement methods in conjunction with EGAIC.
  • To provide insights into the reliability of EGAIC across different metabolic cart configurations.

Main Methods:

  • Six males and 14 females underwent exercise testing on a stationary cycle ergometer.
  • Pulmonary minute ventilation (V̇E) was measured simultaneously with EGAIC computations during warm-up, an incremental ramp test to exhaustion, and a steady-state exercise bout at 80% of the ventilatory threshold.
  • Three airflow sensor methods—turbine (fT), pneumotachograph (fP), and Venturi (fV)—were assessed for their impact on V̇E measurements and EGAIC.

Main Results:

  • Exercise intensity significantly influenced V̇E measurements (p < 0.0001).
  • The specific airflow sensor method (p = 0.97) and its interaction with exercise intensity (p = 0.91) did not significantly affect the measurements.
  • Agreement between airflow methods decreased at power outputs eliciting V̇E up to 150 L/min, indicating reduced precision at higher ventilation rates.

Conclusions:

  • While airflow sensor type does not appear to be a primary driver of error in EGAIC across exercise intensities, the precision of measurements decreases at higher ventilation levels.
  • This reduced precision at higher V̇E may complicate the interpretation of biological variability, training responses, and test-retest reliability.
  • Careful consideration of airflow sensor performance and potential variance across different metabolic cart systems is recommended for accurate EGAIC interpretation.