Differential centiloid scale normalization techniques: comparison between hybrid PET/MRI and independently acquired
Ryo Yamakuni1, Takenobu Murakami2, Naoyuki Ukon3
1Department of Radiology and Nuclear Medicine, School of Medicine, Fukushima Medical University, 1 Hikariga-oka, Fukushima, 960-1295, Japan. yamakuni@fmu.ac.jp.
Annals of Nuclear Medicine
|June 20, 2024
Summary
Centiloid (CL) scales derived from amyloid-β (Aβ) PET imaging show nearly equivalent results whether using simultaneously acquired T1-weighted imaging (T1WI) from PET/MRI or separate PET/CT scans. These findings confirm that both normalization methods yield comparable CL values for clinical studies.
Area of Science:
- Neuroimaging
- Radiochemistry
- Biomarker Analysis
Background:
- Centiloid (CL) scales are crucial for semiquantitative analysis of amyloid-β (Aβ) positron emission tomography (PET) imaging.
- CL values are derived from standardized uptake value ratio (SUVR), necessitating T1-weighted imaging (T1WI) for normalization.
- Two T1WI acquisition methods exist: simultaneous (PET/MRI) and separate (PET+MRI).
Purpose of the Study:
- To investigate the correlation and differences in Centiloid (CL) values derived from amyloid-β (Aβ) PET imaging.
- To compare CL values obtained using T1-weighted imaging (T1WI) from simultaneous PET/MRI versus separate PET+MRI acquisitions for normalization.
Main Methods:
- Analyzed 49 patients who underwent Aβ PET/MRI (using 11C-PiB or 18F-FMM) and additional MRI T1WI within 3 years.
- Normalized PET images using both simultaneous PET/MRI and separate PET+MRI methods.
- Calculated SUVR using standard templates and derived CL values for each tracer.
Main Results:
- Strong linear correlations (R2 > 0.99) were observed between PET/MRI and PET+MRI methods for both tracers and acquisition types.
- Wilcoxon signed-rank tests showed no significant differences in CL values overall (p > 0.05), though some differences were noted for specific acquisition parameters.
- Bland-Altman analysis indicated minimal bias between the two normalization methods.
Conclusions:
- Anatomical standardization using PET/MRI and PET+MRI methods yields nearly equivalent Centiloid (CL) values.
- CL values derived from PET/MRI or PET+MRI normalization are consistent and comparable for clinical research applications.