Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Proton beam spot size and position measurements using a multi-strip ionization chamber
Ye Lin1, Haiqun Zhang1, Shuaizhe Gu2
1Department of Particle Beam Applications, Shanghai Advanced Research Institute, Chinese Academy of Sciences, 201200 Shanghai, China.
A new large-area multi-strip ionization chamber (MSIC) accurately measures proton beam spot size and position. This detector is ideal for quality assurance in proton therapy, ensuring precise treatment delivery.
Area of Science:
- Medical Physics
- Radiation Oncology
- Particle Therapy
Background:
- Proton therapy requires precise beam characterization for effective treatment.
- Accurate measurement of proton beam spot size and position is critical for quality assurance.
Purpose of the Study:
- To develop and characterize a large-area multi-strip ionization chamber (MSIC).
- To enable efficient measurement of proton beam spot size and position in proton therapy.
Main Methods:
- A 420 mm x 320 mm MSIC with 240 vertical and 180 horizontal strips was designed.
- Characterization involved irradiating proton spots across 17 energies (73.5-235 MeV) and comparing with EBT3 film.
- Analysis included beam profiles, spot sizes, positions, measurement stability, and sensitivity.
Main Results:
- The MSIC demonstrated excellent agreement with EBT3 film for spot size (±0.18 mm) and position (±0.17 mm) measurements.
- The detector exhibited good short-term measurement stability and low noise performance.
- Accurate characterization was achieved across the clinical proton beam energy range.
Conclusions:
- The developed large-area MSIC provides efficient and accurate proton beam spot characterization.
- The MSIC is suitable for commissioning and quality assurance in pencil beam scanning proton therapy.
- This technology supports fast quantification of spot size and position for improved treatment delivery.