Circuit-dependent carbon dioxide rebreathing during continuous positive airway pressure
Ludovico Messineo1, Bernard Hete2, Ryan Diesem3
1Division of Sleep and Circadian Disorders, Departments of Medicine and Neurology, Brigham & Women's Hospital & Harvard Medical School, Boston, MA, USA.
Background:
The current standard treatment for obstructive sleep apnea (OSA), continuous positive airway pressure (CPAP), is characterized by a low adherence rate due to various factors including circuit-dependent carbon dioxide (CO2) rebreathing, which can exacerbated by disparate factors, such as low PAP, use of auto-titrating PAP or ramps. However, risk factors for rebreathing are often overlooked or poorly understood in clinical practice. Therefore, our objective was to evaluate the extent of rebreathing occurring with commonly used CPAP masks across varying PAPs, tidal volumes, and respiratory rates.
Methods:
In a bench study, we assessed the rebreathing rate of nine masks interfacing a CPAP with a lung simulator providing different breathing respiratory rates (15 or 20 breaths/min) and tidal volumes (400, 500, 600, 700 and 750 mL). Additionally, a theoretical model was developed to describe the likelihood of CO2 rebreathing from four different masks at various breathing settings.
Results:
Overall, all masks performed worse in situations characterized by low PAPs, high tidal volumes, and high respiratory rates. However, Dreamwear, Nuance, Siesta, Vitera, and particularly V2 masks exhibited greater susceptibility to rebreathing compared to F20, P10, Brevida, and Rio masks for the same variations of PAPs or ventilatory parameters. The mathematical model suggested that the risk of rebreathing for Rio, P10 and Nuance mask is negligible for respiratory rates of 10 breaths/min or below.
Conclusions:
Circuit-dependent CO2 rebreathing can be a common occurrence and warrants careful mask selection upon CPAP therapy initiation for optimal clinical outcomes.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Physiological Control of Respiration
Breathing, a seemingly passive process, is regulated by the respiratory center in the brainstem. This center coordinates the involuntary control of respirations, which means it occurs without conscious effort, ensuring a smooth and uninterrupted pattern.
Regulation of Ventilation
The body maintains ventilation by monitoring levels of carbon dioxide (CO2), oxygen (O2), and hydrogen ion concentration (pH) in the arterial blood. Among these factors, the level of CO2 plays a crucial...
Chemical Factors Affecting Respiration Centers
CO2 has a potent influence on respiration and is strictly regulated....
Acute Respiratory Failure-III
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...