Pulsed DPOAEs in serial measurements : Combined analysis paradigm of simultaneously occurring changes in hearing
Katharina Bader1,2, Dennis Zelle3,4, Anthony W Gummer3
1Klinik für Hals-Nasen-Ohren-Heilkunde, Universitätsklinikum Tübingen, Elfriede-Aulhorn-Straße 5, 72076, Tübingen, Germany. Katharina.Bader@med.uni-tuebingen.de.
Background:
To date, there is no consensus on how to standardize the assessment of ototoxicity in serial measurements. For the diagnosis of damage to the cochlear amplifier, measurement methods are required that have the highest possible test-retest reliability and validity for detecting persistent damage. Estimated distortion-product thresholds (LEDPT) based on short-pulse distortion-product otoacoustic emission (DPOAE) level maps use individually optimal DPOAE stimulus levels and allow reliable quantitative estimation of cochlea-related hearing loss.
Materials And Methods:
Hearing thresholds were estimated objectively using LEDPT and subjectively using modified Békésy tracking audiometry (LTA). Recordings were performed seven times within three months at 14 frequencies (f2 = 1-14 kHz) in 20 ears (PTA4 (0.5-4 kHz) < 20 dB HL). Reconstruction of the DPOAE growth behavior as a function of the stimulus levels L1, L2 was performed on the basis of 21 DPOAE amplitudes. A numerical fit of a nonlinear mathematical function to the three-dimensional DPOAE growth function yielded LEDPT for each stimulus frequency. For the combined analysis, probability distributions of hearing thresholds (LTA, LEDPT), DPOAE levels (LDP), and combinations thereof were determined.
Results:
LTA and LEDPT each exhibited a test-retest reliability with a median of absolute differences (AD) of 3.2 dB and 3.3 dB, respectively. Combining LEDPT, LDP, and LTA into a single parameter yielded a significantly smaller median AD of 2.0 dB.
Conclusion:
It is expected that an analysis paradigm based on a combination of LEDPT, suprathreshold LDP, and fine-structure-reduced LTA would achieve higher test performance (sensitivity and specificity), allowing reliable detection of pathological or regenerative changes in the outer hair cells.
Related Concept Videos
Doppler Effect - II
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...