Automated Midline Shift and Intracranial Pressure Estimation based on Brain CT Images
Published on: April 13, 2013
Comparison of image quality between Deep learning image reconstruction and Iterative reconstruction technique for CT
Obhuli Chandran M1, Saikiran Pendem1, Priya P S2
1Department of Medical Imaging Technology, Manipal College of Health Professions, Manipal Academy of Higher Education, Manipal, Karnataka, 576104, India.
Background:
Non-contrast Computed Tomography (NCCT) plays a pivotal role in assessing central nervous system disorders and is a crucial diagnostic method. Iterative reconstruction (IR) methods have enhanced image quality (IQ) but may result in a blotchy appearance and decreased resolution for subtle contrasts. The deep-learning image reconstruction (DLIR) algorithm, which integrates a convolutional neural network (CNN) into the reconstruction process, generates high-quality images with minimal noise. Hence, the objective of this study was to assess the IQ of the Precise Image (DLIR) and the IR technique (iDose 4) for the NCCT brain.
Methods:
This is a prospective study. Thirty patients who underwent NCCT brain were included. The images were reconstructed using DLIR-standard and iDose 4. Qualitative IQ analysis parameters, such as overall image quality (OQ), subjective image noise (SIN), and artifacts, were measured. Quantitative IQ analysis parameters such as Computed Tomography (CT) attenuation (HU), image noise (IN), posterior fossa index (PFI), signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) in the basal ganglia (BG) and centrum-semiovale (CSO) were measured. Paired t-tests were performed for qualitative and quantitative IQ analyses between the iDose 4 and DLIR-standard. Kappa statistics were used to assess inter-observer agreement for qualitative analysis.
Results:
Quantitative IQ analysis showed significant differences (p<0.05) in IN, SNR, and CNR between the iDose 4 and DLIR-standard at the BG and CSO levels. IN was reduced (41.8-47.6%), SNR (65-82%), and CNR (68-78.8%) were increased with DLIR-standard. PFI was reduced (27.08%) the DLIR-standard. Qualitative IQ analysis showed significant differences (p<0.05) in OQ, SIN, and artifacts between the DLIR standard and iDose 4. The DLIR standard showed higher qualitative IQ scores than the iDose 4.
Conclusion:
DLIR standard yielded superior quantitative and qualitative IQ compared to the IR technique (iDose4). The DLIR-standard significantly reduced the IN and artifacts compared to iDose 4 in the NCCT brain.
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Brain Imaging
These technologies include computerized axial tomography (CAT or CT scans), positron-emission tomography (PET scans), magnetic resonance imaging (MRI), functional magnetic resonance imaging (fMRI), and Transcranial Magnetic...