Task-based image quality assessment of an intraoperative CBCT for spine surgery compared with conventional CT
Robert Vorbau1, Markus Hulthén1, Artur Omar2
1Medical Radiation Physics and Nuclear Medicine, Karolinska University Hospital, Stockholm, Sweden.
Purpose:
To analyze the image quality of a novel, state-of-the art platform for CBCT image-guided spine surgery, focusing particularly on the dose-effectiveness compared with conventional CT (the gold standard for postoperative assessment).
Methods:
The ClarifEye platform (Philips Healthcare) with integrated augmented-reality surgical navigation, has been compared with a GE Revolution CT (GE Healthcare). The 3D spatial resolution (TTF) and noise (NPS) were evaluated considering relevant feature contrasts (200-900 HU) and background noise for differently sized patients (200-300 mm water-equivalent diameter). These measures were used to determine the noise equivalent quanta (NEQ) and observer model detectability.
Results:
The CBCT system exhibited a linear response with 50% TTF at 5.7 cycles/cm (10% TTF at 9.2 cycles/cm), and the axial noise power peaking at about 3.6 cycles/cm (average frequency of 4.1 cycles/cm). The noise magnitude and texture differed markedly compared to iteratively reconstructed CT images (GE ASiR-V). The CBCT system had 26% lower detectability for a high-frequency task (related to edge detection) compared with CT images reconstructed using the Bone kernel combined with ASiR-V 50%. Likewise, it had 18% lower detectability for low- and mid-frequency tasks compared with CT images reconstructed using the Standard kernel. This difference translates to 50%-80% higher CBCT imaging doses required to match the CT image quality.
Conclusions:
The ClarifEye platform demonstrates intraoperative CBCT-imaging capabilities that under certain circumstances are comparable with conventional CT. However, due to limited dose-effectiveness, a trade-off between timeliness and radiation exposure must be considered if end-of-procedure CBCT is to replace postoperative CT.