Evaluating the functionality of x-ray angiography systems based on DICOM structured reports
Artur Omar1, Markus Hulthén2, Angeliki Karambatsakidou3
1Medical Radiation Physics and Nuclear Medicine, Karolinska University Hospital, Stockholm, Sweden; Department of Clinical Science, Intervention and Technology, Karolinska Institutet, Stockholm, Sweden.
Evaluating x-ray angiography system functionality using DICOM radiation dose structured reports (RDSR) is feasible. Detailed exposure data from RDSRs enables system performance assessment, though some analyses require extensive clinical data due to standard limitations.
Area of Science:
- Medical Imaging Physics
- Radiology Informatics
Background:
- The DICOM Radiation Dose Structured Report (RDSR) standard facilitates the collection of detailed exposure data from medical imaging systems.
- Assessing the technical functionality of x-ray angiography (XA) systems is crucial for ensuring diagnostic image quality and patient safety.
Purpose of the Study:
- To determine the feasibility of evaluating XA system functionality using exposure data from DICOM RDSRs.
- To explore applications of RDSR data for system performance monitoring and characterization.
Main Methods:
- Utilized RDSR data from Philips Healthcare XA systems across various clinical suites (cardiology, radiology, neurology).
- Collected data over seven years using a DICOM server and custom parser to analyze irradiation events.
- Developed methods to evaluate XA system performance, including tube output, exposure geometry, and automatic control logic.
Main Results:
- Demonstrated feasibility of evaluating XA system functionality via RDSR data.
- Validated a method for assessing X-ray tube output, showing excellent agreement with conventional measurements and theoretical predictions.
- Established reference data for exposure geometry and beam quality across different angiography procedures.
Conclusions:
- Evaluating XA system technical functionality using RDSRs is feasible.
- Limitations in the current DICOM RDSR standard necessitate substantial clinical data for certain analyses.
- RDSR data offers a valuable resource for system performance monitoring and establishing procedural references.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
X-ray Imaging
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...