Related Experiment Video
Updated: Jun 20, 2025

Standardized Technique of Aortic Valve Re-implantation for Valve-sparing Aortic Root Replacement
Published on: December 11, 2017
Effects of implantation height on the performance of a redo transcatheter aortic valve replacement using a
Huang Chen1, Milad Samaee1, Pradeep Yadav2
1Department of Biomedical Engineering, Georgia Institute of Technology and Emory University, Atlanta, Ga.
Objective:
The use of the transcatheter aortic valve in low-risk patients might lead to a second intervention due to the deterioration of the first 1. Understanding the implantation height is key to an effective redo transcatheter aortic valve replacement treatment.
Methods:
The effects of implantation height on the performance of a balloon-expandable valve within a self-expandable valve were assessed using hemodynamic testing and particle image velocimetry. The hemodynamic performances, leaflet kinematics, and turbulent shear stresses were measured and compared.
Results:
When a second balloon-expandable valve was positioned at varying heights relative to the first self-expandable valve, the leaflet motion of the first valve transitioned from free opening and closing to overhanging, and eventually to being entirely pinned to the stent, forming a neo-skirt. When the leaflets of the self-expandable valve could move freely, a decrease in regurgitation fraction was observed, but with an increased pressure gradient across the valve. Flow visualization indicated that the overhanging leaflets disrupted the flow, generating a higher level of turbulence.
Conclusions:
This study suggests that the overhanging leaflets should be avoided, whereas the other 2 scenarios should be carefully evaluated based on an individual patient's anatomy and the cause of failure of the first valve.
More Related Videos
12:17Full-root Aortic Valve Replacement by Stentless Aortic Xenografts in Patients with Small Aortic Roots
Published on: May 21, 2017
05:31Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022