Optimization of handheld spectrally encoded coherence tomography and reflectometry for point-of-care ophthalmic
Jacob J Watson1, Rachel Hecht1, Yuankai K Tao1
1Vanderbilt University, Department of Biomedical Engineering, Nashville, Tennessee, United States.
Significance:
Handheld optical coherence tomography (HH-OCT) systems enable point-of-care ophthalmic imaging in bedridden, uncooperative, and pediatric patients. Handheld spectrally encoded coherence tomography and reflectometry (HH-SECTR) combines OCT and spectrally encoded reflectometry (SER) to address critical clinical challenges in HH-OCT imaging with real-time en face retinal aiming for OCT volume alignment and volumetric correction of motion artifacts that occur during HH-OCT imaging.
Aim:
We aim to enable robust clinical translation of HH-SECTR and improve clinical ergonomics during point-of-care OCT imaging for ophthalmic diagnostics.
Approach:
HH-SECTR is redesigned with (1) optimized SER optical imaging for en face retinal aiming and retinal tracking for motion correction, (2) a modular aluminum form factor for sustained alignment and probe stability for longitudinal clinical studies, and (3) one-handed photographer-ergonomic motorized focus adjustment.
Results:
We demonstrate an HH-SECTR imaging probe with micron-scale optical-optomechanical stability and use it for in vivo human retinal imaging and volumetric motion correction.
Conclusions:
This research will benefit the clinical translation of HH-SECTR for point-of-care ophthalmic diagnostics.
More Related Videos
Related Concept Videos
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography