Related Experiment Video
Updated: Jun 18, 2025

Automated Acoustic Dispensing for the Serial Dilution of Peptide Agonists in Potency Determination Assays
Published on: November 10, 2016
Development of a sensitive LC-MS/MS assay to support human microdose study for an oral agonist of the GLP-1 receptor
Eugene P Kadar1, Heather Eng1, Amit S Kalgutkar2
1Medicine Design, Pfizer Worldwide Research & Development, Pfizer Inc., Groton, CT 06430, USA.
Abstract:
Aim: The purpose of this work was to determine the feasibility of supporting a clinical microdose study for PF-06882961 (danuglipron), an oral small molecule agonist of the GLP-1 receptor, by LC-MS/MS. Methodology: Statistical instrument parameter optimization using response surface methodology was employed to develop a LC-MS/MS method for the analyte, PF-06882961. Results: An LC-MS/MS method was developed and validated to support a proof of concept microdose pharmacokinetics preclinical study in monkeys, administered PF-06882961 (0.005 mg total, average dose = 0.0007 mg/kg) via intravenous bolus injection. Conclusion: The present study demonstrated the feasibility of analyzing human microdose plasma samples for PF-06882961 by LC-MS/MS, instead of accelerator mass spectrometry, thereby reducing cost and eliminating synthesis and exposure to 14C labeled material.

