Related Experiment Video
Updated: Jul 10, 2026

A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Left ventricular assist device implantation outcomes in patients with subcutaneous implantable
Alex J Nusbickel1, Stephen Allan Petty1, Steven J Ross1
1Division of Cardiovascular Medicine, University of Florida College of Medicine, Gainesville, FL, United States of America.
Background:
Left ventricular assist devices (LVADs) may induce electromagnetic interference (EMI) affecting implanted cardiac devices, including more novel subcutaneous implantable cardiac defibrillators (S-ICDs).
Methods:
In this case series, the authors retrospectively reviewed courses of 6 patients with S-ICDs who underwent LVAD implantation at a single center.
Results:
Of the 6 patients reviewed, 4 experienced inappropriate ICD shocks, of which 3 resulted from EMI. Five of the 6 patients ultimately had S-ICD therapies disabled.
Conclusions:
Due to EMI resulting in inappropriate shocks and improved tolerability of malignant arrhythmias, deactivation or removal of S-ICDs should be considered in patients undergoing LVAD implantation.
Related Concept Videos
Heart Failure VI: Adjunct Therapies
Cardiomyopathy II: Dilated Cardiomyopathy
Cardiomyopathy V: Interprofessional Care