Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

X-ray Imaging01:24

X-ray Imaging

5.4K
German physicist Wilhelm Röntgen (1845–1923) was experimenting with electrical current when he discovered that a mysterious and invisible "ray" would pass through his flesh but leave an outline of his bones on a screen coated with a metal compound. In 1895, Röntgen made the first durable record of the internal parts of a living human: an "X-ray" image (as it came to be called) of his wife’s hand. Scientists worldwide quickly began their own experiments with...
5.4K

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Study of dose reduction in computed tomography using a time-of-flight scatter rejection mechanism in a 3D voxelized phantom.

Physics in medicine and biology·2025
Same author

Study of the TOFPET2c ASIC in time-of-flight detection of x-rays for scatter rejection in medical imaging applications.

Physics in medicine and biology·2025
Same author

Learning to see via epiretinal implant stimulation<i>in silico</i>with model-based deep reinforcement learning.

Biomedical physics & engineering express·2023
Same author

Efficient numerical modelling of time-domain light propagation in curved 3D absorbing and scattering media with finite differences.

Biomedical optics express·2021
Same author

3D Photon-to-Digital Converter for Radiation Instrumentation: Motivation and Future Works.

Sensors (Basel, Switzerland)·2021

Related Experiment Video

Updated: Jun 16, 2025

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
08:30

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging

Published on: September 11, 2011

14.4K

Scatter estimation and correction using time-of-flight and deconvolution in x-ray medical imaging.

Julien Rossignol1,2, Gabriel Bélanger1,2, Maëlle Fromont1,2

  • 1Institut interdisciplinaire d'innovation technologique (3IT), Université de Sherbrooke, Sherbrooke, Québec, Canada.

Physics in Medicine and Biology
|August 15, 2024
PubMed
Summary

A novel iterative deconvolution method significantly improves scatter rejection in time-of-flight (TOF) X-ray imaging. This technique recovers 89% of contrast degradation, outperforming simple thresholding and easing system requirements.

Keywords:
computed tomographydeconvolutionmedical imagingradiographyscatter correctionscatter rejectiontime-of-flight

More Related Videos

Quantifying X-Ray Fluorescence Data Using MAPS
14:58

Quantifying X-Ray Fluorescence Data Using MAPS

Published on: February 17, 2018

10.7K
Sample Drift Correction Following 4D Confocal Time-lapse Imaging
10:04

Sample Drift Correction Following 4D Confocal Time-lapse Imaging

Published on: April 12, 2014

16.4K

Related Experiment Videos

Last Updated: Jun 16, 2025

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
08:30

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging

Published on: September 11, 2011

14.4K
Quantifying X-Ray Fluorescence Data Using MAPS
14:58

Quantifying X-Ray Fluorescence Data Using MAPS

Published on: February 17, 2018

10.7K
Sample Drift Correction Following 4D Confocal Time-lapse Imaging
10:04

Sample Drift Correction Following 4D Confocal Time-lapse Imaging

Published on: April 12, 2014

16.4K

Area of Science:

  • Medical Imaging
  • Photonics
  • Computational Imaging

Background:

  • Time-of-flight (TOF) scatter rejection in X-ray imaging requires stringent timing jitter (≤50 ps) to effectively reduce scattered photons.
  • Existing methods like anti-scatter grids can attenuate primary photons, impacting image quality.
  • The timing jitter contribution from both the X-ray source and detector can be measured and potentially corrected.

Purpose of the Study:

  • To develop and evaluate an iterative deconvolution method for scatter correction in TOF X-ray imaging.
  • To assess the effectiveness of this method in improving image quality across various timing jitter levels.
  • To compare the performance of the deconvolution method against traditional time thresholding techniques.

Main Methods:

  • Simulated radiography and CT imaging using GATE software.
  • A scatter correction algorithm based on iterative deconvolution of the measured time point-spread function.
  • Evaluation using phantoms including a water cylinder with bone inserts and a head and torso model.
  • Testing with total timing jitters ranging from 100 ps to 500 ps full-width-at-half-maximum (FWHM).

Main Results:

  • The deconvolution method recovered 89% of contrast degradation in head and torso radiography at 200 ps FWHM jitter, compared to 28% with simple time thresholding.
  • Corrected images showed a percent root-mean square error between 2% and 14% across all tested jitter levels (100-500 ps FWHM).
  • Performance with the deconvolution method surpassed scatter rejection alone at 100 ps FWHM.

Conclusions:

  • Iterative deconvolution offers a promising approach for scatter correction in TOF X-ray imaging, significantly enhancing image quality.
  • This method demonstrates lower system requirements for timing jitter compared to simple TOF scatter rejection.